![Rohde & Schwarz SMBV100A Скачать руководство пользователя страница 561](http://html1.mh-extra.com/html/rohde-and-schwarz/smbv100a/smbv100a_operating-manual_1477735561.webp)
Remote Control Commands
R&S
®
SMBV100A
546
Operating Manual 1176.8016.02 ─ 17
:CALibration:DATA:FACTory:DATE?
Queries the date of the last factory calibration.
Return values:
<Date>
string
Example:
CAL:DATA:FACT:DATE?
// "2010-01-01"
Usage:
Query only
:CALibration<hw>:FMOFfset[:MEASure]?
Starts all adjustment for the FM/PhiM modulator.
Return values:
<Measure>
0 | 1
Example:
CAL:FMOF?
starts the adjustments for the FM/Phim modulator.
Response: "0"
the adjustments have been performed successfully
Usage:
Query only
Manual operation:
See
:CALibration<hw>:FREQuency[:MEASure]?
Starts all adjustments which affect the frequency.
Return values:
<Measure>
0 | 1
Example:
CAL:FREQ:MEAS?
starts the adjustments for maximum frequency accuracy.
Response: "0"
the adjustments have been performed successfully.
Usage:
Query only
Manual operation:
See
:CALibration<hw>:IQModulator:FULL?
Starts adjustment of the I/Q modulator for the entire frequency range. The I/Q modula-
tor is adjusted with respect to carrier leakage, I/Q imbalance and quadrature.
Return values:
<Full>
0 | 1
*RST:
0
CALibration Subsystem
Содержание SMBV100A
Страница 1: ...R S SMBV100A Vector Signal Generator Operating Manual Operating Manual 1176 8016 02 17 ÚÞ 2 ...
Страница 31: ...Contents R S SMBV100A 16 Operating Manual 1176 8016 02 17 ...
Страница 129: ...Manual Operation R S SMBV100A 114 Operating Manual 1176 8016 02 17 Legend of Front Panel Controls ...