Chapter 4
Using the Capnostream Monitor
Preparing the Monitor for a Patient
Using Patient Cases and Patient ID Numbers
Changing the Alarm and Pulse Volumes
Turning the Pump Off for Suction or Lavage
Monitor Screen Menu Reference Chart
Preparing the Monitor for a Patient
CAUTION:
If any monitor response does not seem appropriate, do not use the monitor. Instead, contact
your local representative.
The following steps describe the procedure for preparing the monitor for a patient.
T
O
P
REPARE THE
M
ONITOR FOR A
P
ATIENT
:
1. Turn on the monitor by pushing the on/off switch
on the front panel.
2. The complete power on sequence is described in
3. Confirm that the green power-on indicator and the orange AC power indicator light are both on.
CAUTION:
If the orange light is not on, the monitor is running on battery power only and will stop
operating when the battery is discharged. See
4. The red alarm and yellow alarm lights will briefly light up and the speaker will beep.
CAUTION:
If the red and yellow alarm lights do not light up or there is no sound from the speaker, the
monitor should not be used and should be sent for servicing.
5. Connect one or both sensors to the monitor, following the instructions in this manual. To connect the
on page
Connecting an SpO2 Sensor to the Monitor
page
Once either or both sensors are connected to the monitor, it is ready for operation.
If a FilterLine is not connected, there is no CO
2
waveform and the message
FILTERLINE DISCONNECTED
will appear.
If an SpO
2
sensor is not connected, there is no SpO
2
waveform and the message
SpO
2
SENSOR
DISCONNECTED
will appear.
It is possible to use either the Capnography function (etCO
2
) or the Pulse Oximetry function (SpO
2
) without
using the other function. If you only want to operate one function, connect ONLY the sensor for that function,
and the monitor will operate normally.
Note:
The Capnostream monitor is for use with one patient at a time only.
Setting the Patient Type
There are five different patient types recognized by the monitor, listed below.
Infant/Neonatal: for patients from birth to the age of one year
Pediatric 1-3 yrs: for patients aged one to three years
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......