Returning the Monitor
118
Portable Bedside Capnograph/Pulse Oximeter
Printer paper is either not threaded
properly over plastic cover or
caught in the plastic cover.
Open the plastic printer cover and pull the
paper so that a short length is outside the
monitor. Hold the paper so that the short
length of printer paper remains outside the
monitor, and then close the cover so that it
clicks into place.
Printer is out of paper.
Open the plastic cover and insert a new roll
of paper.
Printer works, but
output paper is blank.
Paper roll is placed backwards in
the printer compartment of the
monitor.
Open the plastic cover, turn the roll of paper
the other way, and replace the plastic cover,
taking care to leave a short length of paper
outside the monitor.
Nurse Call
Problem
Cause
Action
Nurse Call output does
not work.
Nurse Call not enabled.
Enable the Nurse Call function from the
system setup screen, or from the Institutional
Defaults screen in Service Mode.
Wiring problem on phono plug.
Check the wiring of the cable and phono plug
connected to the Nurse Call socket on the
back of the monitor.
CO
2
Calibration
Problem
Cause
Action
CALIBRATION
REQUIRED
message
appears on the monitor,
but salutation screen
shows there are still
hours remaining before
next calibration.
It has been more than one year since
the last CO
2
calibration.
Perform a CO
2
calibration.
Returning the Monitor
If it is necessary to return the monitor for repairs, contact your local representative for shipping instructions.
To repack the monitor, disconnect the accessories from the monitor. Pack the monitor in the original shipping
carton. If the original carton is unavailable, use a suitable box filled with the appropriate amount of packing
material. It is not necessary to return the sensors, Microstream
®
EtCO
2
consumables, or power cords.
If the monitor malfunctions, carefully package the monitor with an unused consumable from the same box or lot
as the consumable used at the time of malfunction and return it with the monitor for inspection.
Technical Assistance
For technical information, contact your local Oridion representative or write to
The Service Manual includes information that is required by qualified personnel to service the monitor.
See also the Technical Service area in the Capnography section of our website
If it is necessary to return the monitor for repairs, contact your local representative for shipping instructions.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......