Adjustable SpO2 Parameters
62
Portable Bedside Capnograph/Pulse Oximeter
Figure 26 - SpO
2
Section of Numeric Home Screen
Adjustable SpO
2
Parameters
The Capnostream monitor provides the option of adjusting some parameter settings used for SpO
2
measurement
to suit your patients, your institution's requirements, or other needs. To change these settings on a temporary
basis, until the device is turned off, follow the procedure below. To set changes as institutional defaults so that
the settings will remain in effect even after the monitor is turned off, see
on page
T
O
C
HANGE
S
P
O
2
P
ARAMETER
S
ETTINGS
:
1. On the Home screen, move the control knob to the SpO
2
section of the screen, so that it is outlined in blue.
Click the control knob.
2. The SpO
2
Setup screen will appear. Move the control knob to the parameter that you wish to change and
click to select that parameter. Move the control knob to select the desired setting and click to select the
setting. A list of the settings that can be changed appears in
Table 12 - Adjustable SpO2 Parameters
3. Move the control knob to the
HOME
button and click to select. The screen will also revert to the Home
screen after a few seconds if no additional actions are taken on that screen.
4. The changes in the parameters will remain in effect until the device is turned off.
Table 12 - Adjustable SpO
2
Parameters
Parameter
Choices
Factory Default
Pulse Tone
On/Off
On
Sat Sec (Sat Seconds)
On/Off
On
SpO2 Scale for Trend Display
0-100, 50-100
50-100
PR Scale for Trend Display
150, 300
150
Sweep Speed (mm/sec)
[for current patient type]
3, 6.3, 12.5, 25
25
SPO
2
Alarm Limit Message
When the SpO
2
alarm limit is set below 85%, a message reading SpO
2
LOW ALARM LIMIT: xx will appear in
the header area, indicating the level of the SPO
2
LOW alarm limit.
SpO
2
, Sat
Seconds and
Pulse Rate
values
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......