Printed Report Options
Portable Bedside Capnograph/Pulse Oximeter
97
Printed Report Options
Capnostream can be purchased with a built-in thermal strip printer. The report printing menu in Capnostream is
for use with the optional printer.
To print a report on an external printer, the recommended procedure is to transfer the data to a computer using a
USB flash memory device (see
Data Transfer via the USB Data Port
). The report can then be
formatted and printed using the computer.
The following printed reports are available:
•
Tabular Case Report
•
Graphical Case Report
•
Tabular Trend Report
•
Graphical Trend Report
•
Real Time Continuous Waveforms
•
Real Time Continuous CO
2
Waveform
•
Real Time Continuous Tabular
The printed Apnea and O
2
Desat Report (if available) is discussed above.
The data printed for the trend reports is the data that was last displayed on the Trend screen at the time
TREND
was selected. The resolution of the case report is always the minimum resolution (maximum detail).
Real Time Continuous Tabular data is printed at the same interval that the numerics on the screen are updated.
The Real Time Continuous Waveform graph/s are printed as displayed on the screen.
Please note that all Trend and Case reports must be printed
before
the case is ended. Once a case is ended, the
case and trend data is deleted from the memory, and printing is no longer possible.
Printed Reports
The Print screen is accessed from the Apnea and O
2
Desat Report screen (if A/hr and ODI are enabled) or the
Home screen (if A/hr and ODI is not available or not enabled).
The Print screen allows you to choose what report to print, and also to start and stop the printing of a report.
T
O PRINT A REPORT
:
1. From the Home screen, select
REPORTS
.
2. On the Apnea and O
2
Desat screen which will appear, select
to display the Print screen seen in
3. For Infant/Neonatal patients, select
on the Home screen to view the Print screen.
4. Use the control knob to select the type of report to print. Only one type of report can be selected at a time.
An asterisk (*) will indicate the report that has been selected. If you choose a case report and no case is
currently active, the field to the right of the report name will read NO CASES.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......