Operation with Hospital Patient Data Systems
110
Portable Bedside Capnograph/Pulse Oximeter
Medium Priority Alarms
YES
Advisories
NO
Silent Advisories
NO
Operation with Hospital Patient Data Systems
The Capnostream monitor provides connectivity with hospital patient data systems (Bernoulli
®
and Oxinet
®
III)
produced and/or marketed by Cardiopulmonary Corporation (CPC). This option permits regular, real-time
transfer of data from the monitor to hospital patient data systems. Eight-bed or 12-bed configurations are
available.
Before beginning the connection process, ensure that the following equipment is available:
•
Bernoulli
®
or Oxinet
®
III system installed in the hospital
•
Bernoulli-MSM or Oxinet Client Bridge terminated with a 9 pin D connector cable
•
Capnostream monitor
Connect the system as described in the DFU supplied with the Bernoulli-MSM or Oxinet Client Bridge. The
system is used with a Y-cable attaching the client bridge, the Capnostream, and wall power. Ensure that all
connectors are securely attached so that the Capnostream continues to receive wall power during operation with
this system.
Once the connection between the devices is made as described above, data in binary format will be transferred
automatically from the Capnostream monitor to the Bernoulli
®
/Oxinet
®
III system. No additional setting of the
monitor is required.
The following measurement data is transferred:
•
Instantaneous CO
2
•
EtCO
2
•
FiCO
2
•
Resp rate
•
SpO
2
•
Pulse
In addition, information regarding patient type, alarm data, and device settings (alarm limits, etc.) are
transferred.
An outbound vital signs interface to a clinical information system as well as an ADT (admit, discharge &
transfer) inbound interface is also available from the Bernoulli system.
For more information regarding the Bernoulli
®
/Oxinet
®
III system, or for troubleshooting the setup procedure,
contact your local distributor.
CAUTION:
If connectivity with the Hospital Patient Data System fails as a result of a fault in the
Hospital Patient Data System or its cabling, Capnostream will not send data to the system.
CAUTION:
Only turn off the Capnostream audible alarm indication after assuring that the patient will
be continuously monitored by other acceptable means.
Operation with Nuvon VEGA Systems
The Capnostream can also connect with the VEGA system from Nuvon for the purposes of interfacing to
clinical information systems to electronically chart patient vital signs.
Capnostream connection to the VEGA system is completely accomplished by the Nuvon implementation
specialist on site with the assistance of an assigned hospital IT department representative. All physiologic
patient data developed by the Capnostream is sent to the VEGA system to be used by the clinical information
system. For more information regarding the Nuvon VEGA system contact an Oridion sales representative or a
Nuvon sales representative directly.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......