Standard Sections of the Display Screen
Portable Bedside Capnograph/Pulse Oximeter
35
Menu Bar
The menu bar of available options and functions is located along the bottom of the monitor display screen. At
the left hand side is the battery charge level indicator. At the right hand side is the speaker volume control.
The menu bar will change depending on which options and functions are available for a specific screen. In some
screens there are additional selectable options in the screen-specific area.
Real Time CO
2
Data
This area of the screen displays the real time etCO
2
and FiCO
2
values along with selected units, and the
respiration rate (RR) in breaths per minute. Numeric display appears at right and a waveform display at left. For
more details regarding the displayed information, see
CO2 Data Displayed by the Capnostream Monitor
on
When the CO
2
alarm limit threshold is crossed, the device will alarm: the affected numeric will flash, the
numeric will appear with a red or yellow background (depending on whether it is a high priority patient [urgent]
alarm or a caution alarm), and a message will appear in the header area of the screen.
This area of the screen can be selected as if it were a menu option. Selecting this area gives access to the setup
screen for changing the CO
2
parameter values.
Real Time SpO
2
Data
This area of the screen displays real time numeric SpO
2
data: SpO
2
, PR, and the pleth waveform or pulse beat
bar - is always shown. (If IPI is enabled and thus no SpO
2
waveform is displayed, the SpO
2
pulse beat is
represented by the green bar at the left of this box.)
For more details about the displayed SpO
2
data, see
Chapter 6 Pulse Oximetry with the Capnostream Monitor
on page
When the SpO
2
alarm limit threshold is crossed, the device will alarm: the affected numeric will flash, the
numeric will appear with a red or yellow background (depending on whether it is a high priority patient [urgent]
alarm or a caution alarm), and a message will appear in the header area of the screen.
This area of the screen can be selected as if it were a menu option. Selecting this area gives access to the setup
screen for changing the SpO
2
parameter values, including the option to turn the pulse beat tone on or off.
Real Time IPI Data with Trend Graphical Display
This area of the screen displays the real-time value of the IPI, a numerical value which summarizes four
parameters (etCO
2
, RR, SpO
2
, PR) in order to provide a quick indication of the patient’s respiratory status. The
IPI trend graph indicates the trend of this value over the chosen period of time (the default period is one hour).
On the IPI trend graph, the red and yellow horizontal lines indicate the currently set high priority patient
(urgent) alarm and caution alarm limits respectively.
For more information about this parameter, see
Chapter 7 Integrated Pulmonary Index™
When the IPI alarm limit is exceeded, the device will alarm: the affected numeric will flash, the numeric will
appear with a red or yellow background (depending on whether it is a high priority patient [urgent] alarm or a
caution alarm), and a message will appear in the header area of the screen.
IPI data is not available for some patients; for more information, see
Chapter 7 Integrated Pulmonary Index™
A/hr and ODI Data
This area of the screen displays the A/hr and ODI values (if available).
For more information about A/hr and ODI, see
Apneas per Hour and the Oxygen Desaturation Index
page
When the apnea count in any one hour period during the last 12 hours exceeds a set number, a Visual Alert, in
the form of an asterisk which appears next to the A/hr value, will appear on the screen, indicating to the user
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......