Types of Alarms
Portable Bedside Capnograph/Pulse Oximeter
75
Message
Description
Corrective Action
REMOTE SYSTEM
DISCONNECTED
Capnostream is no longer
connected to a remote system.
This message will only be present
if enabled by the host computer,
and may have different wording if
so programmed by the host
computer.
Check connection with remote
system
Advisories
Table 17 - Advisories
Message
Description
CLEARING FILTERLINE
FilterLine kinked or clogged with water. Appears
during clearing time until FilterLine is unclogged, or
a blockage state is determined.
NO USB DEVICE FOUND
A valid flash memory device is not connected to
the USB port.
USB FLASH FULL
No room on the USB flash memory device.
USB TIME OUT
USB communication stopped due to lack of
response from the USB device.
SPO
2
WEAK. REPOSITION SENSOR.
SPO
2
WEAK. TOO MUCH LIGHT.
SPO
2
WEAK. TRY EAR SENSOR.
SPO
2
WEAK. TRY NASAL SENSOR.
SPO
2
WEAK. TRY ADHESIVE SENSOR
SPO
2
WEAK. TRY USING HEADBAND
SPO
2
WEAK. SENSOR TOO COLD.
SPO
2
WEAK. CHECK BANDAGE.
SPO
2
WEAK. NAIL POLISH?
SPO
2
WEAK. SENSOR TOO TIGHT?
SPO
2
WEAK DUE TO INTERFERENCE.
SPO
2
WEAK. CLEAN SENSOR SITE.
SpO
2
module detects
a weak pulse and suggests
possible causes.
Silent Advisories
Table 18 - Silent Advisories
Message
Description
CO
2
WARM UP
CO
2
module is preparing for operation.
CO
2
READY
Before the first measurement of CO
2
, after the FilterLine is
connected and before patient breath is detected,
CO
2
READY
replaces the
CO
2
WARM-UP
message.
CALIBRATION REQUIRED
CO
2
calibration is overdue.
MAINTENANCE REQUIRED
CO
2
maintenance is overdue.
PERFORMING AUTOZERO
The monitor automatically performs a zeropoint calibration.
USB DATA TRANSFER IN
PROGRESS
USB data communication in progress.
RS-232 DATA TRANSFER IN
PROGRESS*
RS-232 data communication in progress.
USB DATA TRANSFER ABORTED
USB data communication has been stopped.
RS-232 DATA TRANSFER
ABORTED
RS-232 data communication has been stopped.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......