Replacing the Printer Paper Roll
Portable Bedside Capnograph/Pulse Oximeter
115
Replacing the Printer Paper Roll
If the printer runs out of paper, replace it with a roll of thermal printer paper (Oridion part # 010516) or similar
paper which meets the specifications outlined in the specifications in
Internal Thermal Printer (optional)
page
T
O REPLACE THE PRINTER PAPER
:
1. Open the plastic cover on the printer.
2. Remove the empty spindle inside the paper compartment.
3. Insert a new roll of paper in the direction shown in the figure below, so that the loose end of the paper comes
out at the top of the plastic cover as shown.
Figure 45 - Insert Paper Roll into printer
4. Close the door so that it clicks shut. Briefly press the Feed button to verify that the paper is aligned properly
and is not caught on the edge of the cover.
Cleaning
To clean the monitor’s surfaces, lightly dampen a cloth with a 70% alcohol solution and wipe all surfaces.
Alcohol wipes may also be used. Frequency of the cleaning procedure should be in keeping with hospital
policy.
To clean the screen, use a damp, lint-free cloth.
To clean Nellcor reusable sensors, first remove the sensor from the patient and disconnect it from the
Capnostream device. The sensor may be surface-cleaned with a solution such as 70% isopropyl alcohol. If low-
level disinfection is required, use a 1:10 bleach solution. Do not use undiluted bleach (5%~5.25% sodium
hypochlorite) or any cleaning solution other than those recommended here, because permanent damage to the
sensor could occur. The ear clip sensor may be cleaned by wiping or soaking it (for 10 minutes) in isopropyl
alcohol (70%). If the ear clip is soaked, be sure to rinse it with water and air dry it prior to use on the next
patient. Do not sterilize the ear clip by irradiation, steam, or ethylene oxide. Please refer to the DFU for each
reusable sensor for specific instructions.
WARNING:
Do not autoclave or sterilize this device.
Feed Button
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......