Terminating Operation of the Monitor
Portable Bedside Capnograph/Pulse Oximeter
37
When the etCO
2
, FiCO
2
, or RR alarm limit threshold is crossed, the device will alarm: the affected numeric will
flash, the numeric will appear with a red or yellow background (depending on whether it is a high priority
patient urgent alarm or a caution alarm), and a message will appear in the header area of the screen.
This area of the screen can be selected as if it were a menu option. Selecting this area gives access to the setup
screen for changing the CO
2
parameter values.
Real Time SpO
2
Data
This area of the screen displays real time SpO
2
data. Numeric display is provided, sized so that it can easily be
read even at a distance. As in the standard screen, only the parameters which are measured by the probe which
is currently connected to the device will appear on the screen. For more details about the displayed SpO
2
data,
see
Chapter 6 Pulse Oximetry with the Capnostream Monitor
When the SpO
2
alarm limit threshold is crossed, the device will alarm: the affected numeric will flash, the
numeric will appear with a red or yellow background (depending on whether it is an urgent alarm or a caution
alarm), and a message will appear in the header area of the screen.
A/hr and ODI Data
This area of the screen displays the A/hr and ODI values (if available).
For more information about A/hr and ODI, see
Apneas per Hour and the Oxygen Desaturation Index
on page
When the apnea count in any one hour period during the last 12 hours exceeds a set number, a Visual Alert, in
the form of an asterisk which appears next to the A/hr value, will appear on the screen, indicating to the user
that the A/hr and O
2
Desat screen should be viewed. The A/hr and O
2
Desat screen is reached with the
REPORTS menu button. For more information about the A/hr Visual Alert, see
Real Time IPI Data with Trend Graphical Display
This area of the screen displays the real-time value of the IPI, a numerical value which summarizes four
parameters (etCO
2
, RR, SpO
2
, PR) in order to provide a quick indication of the patient’s respiratory status. The
IPI trend graph below the numeric value indicates the trend of this value over the chosen period of time.
On the IPI trend graph, the red and yellow horizontal lines indicate the currently set urgent alarm and caution
alarm limits respectively.
For more information about this parameter, see
Chapter 7 Integrated Pulmonary Index™
on page
When the IPI alarm limit is exceeded, the device will alarm: the affected numeric will flash, the numeric will
appear with a red or yellow background (depending on whether it is a high priority patient [urgent] alarm or a
caution alarm), and a message will appear in the header area of the screen.
Terminating Operation of the Monitor
Following patient monitoring, operation of the monitor shall be terminated as follows:
1. Remove the FilterLine and SpO
2
sensor from the patient.
2. Switch the monitor to OFF by pressing the ON/OFF switch at the front of the monitor.
3. If the monitor is connected to AC power using the AC adapter, it may now be disconnected from AC power.
To continue charging the monitor when not in use for measurement, the monitor shall remain connected to
AC power. The monitor will continue charging its internal battery even when it is switched off.
Screen Navigation
Select options and set values using the control knob. The control knob works in a manner similar to that of a
regular computer mouse. It is also used as a keyboard for entering letters and numbers.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......