Chapter 13
Maintenance and Troubleshooting
Determining Monitor Service Hours
Replacing the Printer Paper Roll
Introduction
Capnostream requires no routine service other than any performance testing mandated by the operator’s
institution. The monitor requires servicing by qualified service personnel only once every 30,000 operating
hours.
The monitor's CO
2
detection mechanism should be periodically calibrated as detailed below in
. CO
2
calibration can be checked at any time to ensure the calibration is within proper operating
limits.
discusses potential difficulties, possible causes and suggestions for resolving
them.
Note:
Contact your local distributor or refer to the Service Manual for service instructions and performance
tests and checks.
Determining Monitor Service Hours
The information on the Service screen gives the number of hours remaining until servicing or calibration is
required. To access the Service screen, select
SYSTEM
from the Menu bar at the bottom of the Home screen,
and then select
SERVICE
. No password is required to see the number of hours before servicing is required. The
main service screen is shown below.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......