Installing the Battery Pack
24
Portable Bedside Capnograph/Pulse Oximeter
Storing the Battery
The battery pack must be stored in a cold, dry area, not inside the monitor. Its charge decreases over time. To
restore the battery pack to full power, recharge the battery before use. The battery should be fully recharged
every 3 months at minimum. Store at -20 to 25°C.
CAUTION:
Storage or transport of the monitor under environmental conditions beyond those mentioned
in the specification will affect monitor performance and damage the battery and/or the
monitor.
Disposing of the Battery
CAUTION:
Do not dispose of the battery pack in fire; it may explode.
Follow local governing ordinances and recycling instructions regarding disposal or recycling of batteries.
Battery and Power Usage
If power is lost when the monitor is operating from AC power, it automatically switches to the internal battery
pack for power. The duration of this backup power usage, based onbattery capacity, is up to 2.5 hours. The
monitor maintains its settings, including alarm settings, while on battery operation. If settings have been set in
Institutional Defaults, these settings will remain in the monitor memory even if the monitor is not receiving
power at all, and will be available once the monitor is turned on again. Likewise, trend data in the device will
remain in the monitor memory even if the monitor is not receiving power at all, and will be available once the
monitor is turned on again.
The orange AC power indicator light is on when the monitor operates from an external power source, with no
relation to the status of the battery pack.
The green power-on indicator is on when the monitor is switched on.
If the orange AC power indicator light is off and the green power-on indicator is on, the monitor is operating
from the battery pack.
The battery icon will show the battery pack’s approximate charge level. An advisory message,
BATTERY
LOW
, appears when approximately 15 minutes of battery charge (equivalent to 14.0 V) remains.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......