Portable Bedside Capnograph/Pulse Oximeter
11
Do not touch this field
- it is invisible and does not appear in the
final document
Safety Information
To use the Capnostream
®
20
P
monitor (henceforth referred to as Capnostream) correctly and safely, carefully
read this operator’s manual and the
Directions for Use
that accompany Microstream
®
etCO
2
consumables
(FilterLines
®
, henceforth referred to as FilterLines) and the SpO
2
sensors. Use of the monitor
requires full
understanding and strict observance of these instructions, the precautionary information in boldface type, and
the specifications.
Warnings
General
WARNING:
If uncertain about the accuracy of any measurement, first check the patient’s vital signs by
alternate means, and then make sure the monitor is functioning correctly.
WARNING:
The device should not be used as an apnea monitor.
WARNING:
The device should be considered an early warning device. As a trend towards patient
deoxygenation is indicated, blood samples should be analyzed by a laboratory co-oximeter
to completely understand the patient's condition.
WARNING:
To ensure patient safety, do not place the monitor in any position that might cause it to fall
on the patient.
WARNING:
Carefully route patient cabling (SpO
2
sensor and FilterLine) to reduce the possibility of
patient entanglement or strangulation.
WARNING:
Do not lift the monitor by the SpO
2
sensor cable or FilterLine, as they could disconnect
from the monitor, causing the monitor to fall on the patient.
WARNING:
The monitor should not be used adjacent to or stacked with other equipment; if adjacent or
stacked use is necessary, the monitor shall be observed to verify normal operation in the
configuration in which it will be used.
WARNING:
To ensure accurate performance and prevent device failure, do not expose the monitor to
extreme moisture, such as rain.
WARNING:
The use of accessories, transducers, sensors and cables other than those specified may result
in increased emission and/or decreased immunity of the equipment and/or system.
WARNING:
Re-use of single-use accessories could pose a cross-contamination risk to the patient or
damage the functioning of the monitor.
WARNING:
CO
2
readings, respiratory rate, pulse oximetry readings, and pulse signals can be affected by
sensor application errors, certain ambient environmental conditions, and certain patient
conditions.
WARNING:
The monitor is a prescription device and is to be operated by qualified healthcare personnel
only.
WARNING:
No modification of this equipment is allowed.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......