Compliance
Portable Bedside Capnograph/Pulse Oximeter
135
Table 34 - Recommended Separation Distances between Portable and
Mobile RF Communications Equipment and the Monitor
Rated maximum output power
of transmitter W
Separation distance according to frequency
f transmitter in m
1
150 kHz to 80 MHz
d= 1.2
√𝑃
80 MHz to 800 MHz
d= 1.2
√𝑃
800 MHz to 2,5 GHz
d= 2.3
√𝑃
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.27
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures,
objects and people.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......