Types of Nurse Call Systems
Portable Bedside Capnograph/Pulse Oximeter
109
Figure 43 - Connection Point for Nurse Call
Activating Nurse Call
The factory default setting for Nurse Call is disabled, and to operate the feature it must be enabled. This can be
done using the System Setup screen, however, it will reset to disabled again when the monitor is turned off. To
permanently enable the Nurse Call feature, enable it using the monitor screen in the Institutional Defaults
section of the Service mode as follows:
T
O
A
CTIVATE
N
URSE
C
ALL
:
1. Turn on the monitor and wait for the Home screen to appear. Use the control knob to select the
SYSTEM
button to open the System screen, then select
SERVICE
and enter the service password (see
).
2. From the SERVICE screen, select
INST DEFAULTS
(Institutional Defaults) and then select
MONITOR
.
3. Use the control knob to select
NURSE CALL
, and change the option to
ENABLED
.
4. Select
BACK
,
BACK
and
HOME
to exit the Service mode. The new setting is now stored.
5. Test the Nurse Call system as described in
6. Please note that Nurse Call data will not be transferred if alarms have been silenced temporarily (by the
Alarm Silence button at the front of the monitor). See
for details regarding this
feature. Nurse Call data will still be transferred if alarms are silenced through Institutional Defaults (see
on page
Testing Nurse Call
Verify that the system is functioning by forcing a test alarm occurrence (such as breathing into the FilterLine for
a few seconds, then stop breathing into it to create a NO BREATH alarm). Confirm that the expected result was
received according to the standard for the institution's nurse call system. This may be a warning light turned on
or an audio signal generated when the alarm event occurs.
The following table describes which alarms are indicated by the Nurse Call output.
Table 28 - Nurse Call Indicators
Alarm Type
Activates Nurse Call
High Priority (Red) Urgent Patient Alarms
YES
(Yellow) Caution Alarms
NO
Nurse Call
Connection
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......