Apnea and O2 Desaturation Report
Portable Bedside Capnograph/Pulse Oximeter
95
Figure 35 - Apnea and Desat Report Screen
Patient Information
Seen on Screen
Explanation
A/hr
(Apneas per Hour)
Average A/hr per hour over entire time period which
appears on the screen
Apnea Count
Sum of all instances of No breath (apnea) in each time
category (≥10 seconds, 10-19 seconds, 20-29 seconds)
over the entire time period which appears on the screen
SUM(Apnea sum)
Number of instances of no breath (apnea) in each time
category (≥10 seconds, 10-19 seconds, 20-29 seconds) in
each of the intervals within the relevant time period (in this
case, 12:00 AM to 1:00 AM)
Desat Count
Sum of all instances of Desat over the entire time period
which appears on the screen
SUM(Desat sum)
Number of instances of Desaturation in each of the
intervals within the relevant time period (in this case, 12:00
AM to 1:00 AM)
ODI
Average ODI per hour over entire time period which
appears on the screen
Average
Apnea Per
Hour
Desat
Count
Desat Sum
ODI
Apnea
Counts and
Apnea
Sums
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......