Data Transfer via the USB Data Port
106
Portable Bedside Capnograph/Pulse Oximeter
Tabular Trend: 1 hour @30s resolution: 24kB
Real time Continuous CO
2
Waveform: 1 hour @50ms resolution: 4.2MB
Real time Continuous Tabular: 1 hour @2s resolution: 264kB
Full Continuous Transfer: 1 hour @50ms resolution: 12MB
Full Binary Continuous Transfer: 1 hour @50ms resolution: 732kB
Full Binary Trend Transfer: 1 hour @30s resolution: 5kB
USB File Naming Convention
For the different report types, the following file naming convention is used.
<REPORT TYPE>_<PATIENT TYPE >_<REPORT DATE >_<REPORT TIME>_<PATIENT ID >_<FILE NO>.ext
Where:
•
REPORT TYPE - three-letter report type identifier (see
Table 26 - File Naming Conventions
).
•
PATIENT TYPE – patient type (Adult, Pediatric 6-12 yrs, etc).
•
REPORT DATE - Start date on which the report was made in yymmdd format.
•
REPORT TIME - Start time at which the report was made in hhmmss format.
•
PATIENT ID – Patient ID as inputted into the device (or automatically supplied by the device).
•
FILE NO - a running serial number that indicates if the data was split into multiple files.
•
File extension .ext is .txt (tab delimited file type) or .bin (binary file type).
The Patient ID field in the monitor contains the '/' character, which is not a valid character for file names. It is
replaced with a hyphen '-' for file names. The '/' character is used by the monitor to indicate multiple files with
the same Patient ID (e.g.- Smith/1, Smith/2, Smith/3).
File Name Examples:
For different reports taken on January 15, 2011 at 5:23:57 for the patient with Patient ID "20090115035705",
the file names would be:
Table 26 - File Naming Conventions
Report type
File name
Tabular Case Report
TCR_ADULT_110115_052357_20110115035705_1.txt
Tabular Trend Report
TTR_ ADULT _110115_052357_20110115035705_1.txt
Real-time Continuous CO
2
Waveform
RCW_ ADULT_110115_052357_20110115035705_1.txt
Real-time Continuous Tabular
RCT_ ADULT _110115_052357_20110115035705_1.txt
Full Continuous Transfer
FCTR_ ADULT _110115_05235720110115035705_1.txt
Full Binary Continuous Transfer
FCT_ ADULT _110115_052357_20110115035705_1.bin
Full Binary Trend Transfer
FTT_ ADULT _110115_052357_20110115035705_1.bin
Apnea and O
2
Desat Report (if available)
ADR_ ADULT _110115_052357_20110115035705_1.txt
Examples
Taking the same example described above, multiple files for the same Real-time Continuous Tabular report
would look like this:
Report type
File name
Real-time Continuous Tabular
RCT_110115_052357_1.txt RCT_110115_052357_2.txt
RCT_110115_052357_3.txt RCT_110115_052357_4.txt ...
RCT_110115_052357_10.txt ... RCT_110115_052357_100.txt ...
RCT_110115_052357_1000.txt
Note:
Binary files are never split into multiple files because they do not have the length limitation imposed by
MS Excel.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......