Smart A/hr and ODI Home Screen Display
Portable Bedside Capnograph/Pulse Oximeter
67
Note:
Please note that the patient type is used to calculate the A/hr and ODI. For this reason, it is important to
choose the patient type correctly. For the same reason, changing the patient type (from adult to
pediatric, for example) will clear the A/hr and ODI data stored for the current patient. A/hr and ODI
headings will not appear for infant/neonatal or pediatric patients.
Note:
No A/hr or ODI data will appear on the Home screen for the first hour of monitoring, since A/hr and ODI
indicate the number of events over the last hour. Once one hour of monitoring has passed, they will
appear on the Home screen.
Note:
When using the Capnostream A/hr and ODI with a sleeping patient, it is recommended that the
Capnostream device be attached to a central monitoring station where alarms will be heard. Once this is
done, the alarm sound on the Capnostream at the patient’s bedside can be disabled, so as not to disturb
the patient’s sleep. The audible alarms are silenced through
SYSTEM>SERVICE>Service password
(see
on page
>INST DEFAULTS>MONITOR
. In the list of options
on this screen, change the
AUDIO ALARM VOLUME
to
AUDIO OFF
. This should only be done if the
Capnostream is under constant surveillance via connection to a central station (or another means of
surveillance), so that patient alarms are noted by caregivers while the alarm sound is turned off on the
bedside Capnostream device.
Smart A/hr and ODI Home Screen Display
A/hr and ODI appear on the Home screen and are available in all Capnostream functions as a default option,
along with other patient parameters such as EtCO
2
or SpO
2
.
The A/hr and ODI option may be disabled from the
SYSTEM
screen or from the Institutional Defaults screen;
see below for more information. For Infant/Neonatal and Pediatric patients, the Smart A/hr and ODI option is
automatically disabled.
A/hr and ODI Option
T
O
C
HANGE
A/
HR AND
ODI
O
PTION
S
ETTINGS
:
1. On the Home screen, click the
SYSTEM
button in the menu bar at the bottom of the screen.
2. The System Setup screen will appear. Change
A/hr, ODI DISPLAY
to
ENABLED
or
DISABLED
as
desired.
3. Move the control knob to the
HOME
button and click to select. The screen will also revert to the Home
screen after a few seconds if no additional actions are taken on that screen.
4. The changes in the A/hr and ODI display will remain in effect until the device is turned off.
To change the A/hr and ODI option on a more permanent basis, use the Institutional Defaults option, described
in
A/hr and ODI Demo Mode
To view the A/hr and ODI display without connecting a patient, you may set the monitor to Demo Mode by
clicking
System>Service>Enter Service Password>Demo Mode
. (For the Service Password, see
on page
.)
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......