Compliance
134
Portable Bedside Capnograph/Pulse Oximeter
Immunity Test
IEC 60601
test level
Compliance
level
Electromagnetic Environment –
Guidance
Conducted
RF IEC 61000-4-6
3 Vrms 150 kHz to
80 MHz
3 V
Portable and mobile RF
communications equipment should be
used no closer to any part of the
Capnostream20P, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distances:
d= 1.2
√𝑃
d= 1.2
√𝑃
80 MHz to 800 MHz:
d= 2.3
√𝑃
800 MHz to 2.5 GHz:
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation distance in
metres (m).
2
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
ab
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Radiated
RF IEC 61000-4-3
3 V/m 80 MHz to
2.5 GHz
3 V/m
1
U
T
is the a.c. mains voltage prior to application of the test level.
2
At 80 MHz and 800MHz, the higher frequency range applies
NOTE: These guidelines may not apply in all situations. Electronmagnetic propogation is affected by absorption and reflections from
structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the Capnostream20P is used exceeds the applicable RF compliance level above, the
Capnostream20P should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Capnostream20P.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V/m.
The monitor is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the monitor can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the monitor as recommended below, according to the maximum output power of the communications
equipment.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......