Setting Date, Time, Language, and Other Options
38
Portable Bedside Capnograph/Pulse Oximeter
T
O MOVE AROUND THE SCREEN
:
1. Turn the control knob to the right or left to move to the next area on the screen, which is highlighted by the
frame changing to blue.
2. To make a selection, push the control knob in until it clicks.
T
O ENTER LETTERS AND NUMBERS
:
1. When required to enter alphanumeric data, turn the control knob for navigation and then click on the input
data block on the screen.
The content of the data block is cleared and a small block with a yellow background will appear in place of
the first character. In the block will be a black letter A.
2. As you turn the control knob to the right or left, you will see the complete alphabet and numbers 0-9, the
←
backspace symbol, a blank space and the
Enter symbol.
3. When the desired character is displayed, press on the control knob until it clicks. . The inputted character
will return to its regular color (white) and the yellow block will move to the next character space.
4. Turn the control knob again to begin a new character selection.
5. To end alphanumeric entry, press the control knob twice after entering the final letter or symbol.
T
O CHANGE SETTINGS
:
1. When you are on a screen with settings that can be changed, turn the control knob until the desired setting is
highlighted by the frame changing to blue.
2. To select the setting, push the control knob in until it clicks.
3. Turn the control knob until the new setting is reached.
4. To set the setting, push the control knob in until it clicks.
5. Continue to select settings, or turn the control knob to select
BACK
or
HOME
.
Configuration Changes
On most selection screens, when you make a change to one or more system parameters the new settings will
only become effective when you exit the screen by selecting
HOME
or
BACK
. If the selection screen is exited
by pressing one of the front panel control buttons located below the screen, or if you wait more than one minute
and the screen times out and resets, the changed settings will not be saved.
Setting Date, Time, Language, and Other Options
After you first turn on the monitor, check the top left hand corner of the header to ensure that the date and time
are correct.
T
O CHANGE DATE
,
TIME
,
LANGUAGE
,
OR
O
THER
O
PTIONS
:
1. From the Home screen, use the control knob to select
SYSTEM
in the Menu Bar.
2. To adjust the date and time, use the control knob to select and change each setting. The updated date and
time are saved in the monitor memory when you exit the screen, and will not need to be re-adjusted after the
monitor is turned off.
3. To change the language of the display, use the control knob to change the language.
4. The newly set options (including language) will remain in effect until the monitor is turned off. To
permanently change the display language and other settings, see the information in
on page
5. To change any of the other listed parameters, choose the required option, and then select the
HOME
menu
button to return to the Home screen.
6. If A/hr and ODI are not available, the display option will not appear on this screen.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......