Resetting to Factory Defaults
120
Portable Bedside Capnograph/Pulse Oximeter
Resetting to Factory Defaults
Each section of settings described below allows you to reset to the factory settings for that specific section. You
can also do a global reset of all settings in all sections to their factory defaults. To do this, select
RESET
when
you first select the Institutional Defaults screen.
Figure 46 - Institutional Defaults Screen
Uploading or Downloading Institutional Defaults
Capnostream provides the option of uploading institutional defaults from the monitor to a USB Flash Memory
drive or downloading institutional defaults from a USB Flash Memory drive to the monitor.
This utility is used, for example, to simplify the process when an institution wants all devices in a particular
group or section to use the same institutional defaults. In such a case, changes can be made to institutional
defaults on one device, and those defaults can be uploaded to a USB Flash Memory device as described below.
Then, those defaults can be downloaded to the other Capnostream devices in the section/institution using this
procedure.
T
O
U
PLOAD OR
D
OWNLOAD
D
EFAULTS
:
1. For upload, insert a USB Flash Memory drive with at least 2 KB in free memory into the USB port on
Capnostream. For download, insert a USB Flash Memory drive with Institutional Defaults previously loaded
on it into the USB port on Capnostream.
2. Click the
SYSTEM
menu button to enter the main System screen.
3. Click the
SERVICE
menu button to enter the Service screen and input the Service password (see
on page
).
4. Click the
SOFTWARE SUPPORT
menu button to enter the Software Support screen. The Software
Support screen can be seen in
Figure 47 - Software Support Screen
5. Click the
UPLOAD DEFAULTS
button to upload defaults to a USB Flash Memory drive from
Capnostream, or click the
DOWNLOAD DEFAULTS
button to download defaults from a USB Flash
Memory drive to Capnostream. Downloading defaults will create a folder called
settings
on the USB Flash
Memory drive, which will contain files that can be used to upload defaults to another Capnostream. After
uploading defaults, the monitor will shut itself off; restart the monitor to resume operation.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......