Changing Monitor Settings
122
Portable Bedside Capnograph/Pulse Oximeter
Figure 48 - Institutional Defaults Alarm Limits Screen
The factory default settings for Adult/Pediatric and Infant/Neonatal alarm limits are given below.
Table 29 - Factory Default Alarm/Indicator Limits
Parameter
Adult/Pediatric
Red Urgent
Alarm
Adult/Pediatric
Yellow Caution
Alarm
Infant/Neonatal
Red Urgent
Alarm
Infant/Neonatal
Yellow Caution
Alarm
Range
EtCO
2
High
60
55
50
45
5-150 mmHg
EtCO
2
Low
15
17
20
22
0-145 mmHg
FiCO
2
High
8
8
5
5
2-98 mmHg
RR High
30
25
65
60
5-150 bpm
RR Low
5
7
25
28
0-145 bpm
No Breath
Detected
30
N/A
15
N/A
10-60 sec
SpO
2
High
100
100
98
98
25-100%
saturation
SpO
2
Low
85
90
85
90
20-95% saturation
Pulse Rate High
140
140
200
200
30-240 bpm
Pulse Rate Low
50
50
100
100
25-235 bpm
IPI Low Alert
3
5
N/A
N/A
1-9 or OFF
SatSeconds
25
25
25
25
10, 25, 50, 100 or
Off