Data Transfer via the RS-232 Port
Portable Bedside Capnograph/Pulse Oximeter
107
USB Error Messages
The following messages may appear in the message area of the monitor:
NO USB DEVICE FOUND
: Advisory displayed if a USB operation is tried in the absence of a USB device.
USB DEVICE FAILED
: Displayed when the USB device has been detected, but the data transfer operation
cannot be successfully completed.
USB FLASH FULL
: Advisory displayed when data can no longer be downloaded into a USB memory stick
due to the memory being full.
USB TIME OUT
: Advisory displayed when monitor is unable to establish communication with USB device.
Reading Patient Data from Saved Capnostream Files
The USB report types that have a .txt (tab delimited) file type are text files. This makes them easy to read in
most spreadsheet and database software applications. In this case, the .txt format type means that there is a tab
in between each piece of data in every line of the file. The
Patient Data Transfer Application Note
, explaining
the utilization of the transferred data, can be found on the Operator's Manual CD.
The two USB report types that have a .bin file type are binary files. These files are intended for use by
programmers who are creating application programs for use with the Capnostream. The binary file formats are
described in the document
Capnostream Data Transfer Protocols
that can be found on the Operator's Manual
CD.
Data Transfer via the RS-232 Port
Capnostream is equipped with a 9-pin RS-232 connection on the back of the monitor. For detailed information
on the use of this feature, please refer to the document
Capnostream Data Transfer Protocols
, available on the
Operator's Manual CD. The data transfer rate for the RS-232 interface is set in the Institutional Defaults:
Monitor screen. The factory default is automatic detection of the data transfer speed. See
on page
for information on how to change the data transfer rate.
Data transfer via RS-232 can take place at the same time as data transfer with a USB Flash Memory drive.
Note:
The RS-232 port has electrical isolation according to IEC 60101-1-1. Non-medical devices such as PCs
and printers may be connected to this port without additional electrical isolation. These devices must be
placed at least 1.5 meters from the patient environment.
The RS-232 port can be used for data transfer to a PC using Profox. For more details, contact Profox Associates,
Inc. at
Analog Data Output with Capnostream
A digital/analog connection feature is no longer available with Capnostream. Please disregard the digital/analog
(D/A) screens in the device software.
Nurse Call Operation
The Capnostream monitor allows connection to an external Nurse Call system. When connected, the monitor
sends information to the institutional nurse call system that an alarm condition has occurred, alerting medical
personnel that the patient requires medical care. The only data transferred to a Nurse Call system is alarm data,
as described in
Table 28 - Nurse Call Indicators
on page
The nurse call alarm output becomes active simultaneously with the occurrence of an alarm on the monitor, and
remains active while the alarm condition is present. When the alarm condition is no longer present (that is, when
the alarm on the monitor ceases) the Nurse Call alarm output also becomes inactive.
A Nurse Call cable (3.5 m) can be purchased from Oridion (part number 011149). One end of the Nurse Call
Cable attaches to the Capnostream monitor. The cable is supplied un-terminated so it can be built to fit your
nurse call system.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......