Alarm Display
70
Portable Bedside Capnograph/Pulse Oximeter
Alarm Type
Indicators
Audible
Numerics
Messages
Indicator Light
Medium Priority
Repeated triple
beep
N/A
Message area
(some messages
also in waveform
area – see below)
Flashing Yellow
Advisories
Single Beep
N/A
Message area
N/A
Silent Advisories
None
N/A
Message area
N/A
Some messages are displayed in the waveform area as well as in the message area; these messages are:
CO
2
Waveform Area Messages:
•
FILTERLINE BLOCKAGE
•
PERFORMING AUTO ZERO
•
CLEARING FILTER LINE
•
CO
2
ERROR
•
CO
2
STANDBY
•
FILTERLINE DISCONNECTED
SpO
2
Waveform Area Messages:
•
SpO
2
SENSOR NOT ON PATIENT
•
SpO
2
STANDBY
•
SpO
2
SENSOR DISCONNECTED
•
SPO
2
ERROR
•
PULSE NOT FOUND
Alarm Display
In order to view the visual alarm indicators, the intended position of the operator is in front of the monitor
display screen. The intended position of the operator in order to hear audible signals is any position near the
monitor (in a 360° range).
Alarm occurrences are prominently displayed in the real-time numeric section of all screens by the flashing of
the numeric and a change of color of the numeric background: the background of the numeric will flash red if
the value crosses the high priority patient (urgent) alarm limit threshold and flash yellow if the value crosses the
caution alarm limit threshold. (The latter is relevant only if caution alarm limits, which are optional, have been
enabled.)
Alarm occurrences are also displayed on all screens in the header area, in order to provide immediate input on
alarms to the health care provider. The preference of alarm display in the header area is described in
In addition, the message "NO BREATH" and "FILTERLINE BLOCKED" appear in the CO
2
waveform area as
well as in the header if this alarm condition exists. If the numeric home screen option is displayed (so that there
is no waveform area), this message will appear only in the header area.
Likewise, the message "SpO
2
SENSOR NOT ON PATIENT" appears in the SpO
2
waveform area as well as in
the header if this alarm condition exists. If the IPI option is enabled or the numeric home screen is displayed, (so
that there is no waveform area), this message will appear only in the header area.
The alarm occurrences for a specific patient over a specific period of time can be reviewed in the trend screen
(see
Chapter 10 Using Trends
on page
for more details). Capnostream also provides an Alarm Review screen
which displays the absolute number of each type of alarm generated over the last hour for the patient currently
being monitored. The Alarm Review screen is seen in
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......