Monitoring CO2 during MRI Scanning
56
Portable Bedside Capnograph/Pulse Oximeter
4. The changes in the parameters will remain in effect until the device is turned off.
Table 10 - Adjustable CO
2
Parameters
Parameter
Choices
Factory Default
BTPS*
On/Off
On
FiCO
2
Display
On/Off
On
Pump-Off Timeout (minutes)
5, 10, 15 or 30
15
CO
2
Waveform Scale (mmHg)
50, 100, 150, Auto
Auto
EtCO
2
Scale for Trend Display
50, 100, 150
50
RR Scale for Trend Display
50, 100, 150
50
Sweep Speed (mm/sec) [for current patient type]
1, 2, 3, 6.3, 12.5, 25
6.3
A/hr Visual Alert Level (Apnea/Hr) (if available)
1-99
10
* BTPS denotes the standard correction used during measurement for body temperature, pressure, and saturation.
BTPS should be set to ON during all measurement procedures. The device automatically turns off the BTPS
correction during calibration procedures and turns it on again following these procedures. There is no need for the
user to make any changes to the BTPS setting.
Monitoring CO
2
during MRI Scanning
WARNING:
Do not use the FilterLine H Set Infant/Neonatal during magnetic resonance imaging (MRI)
scanning. Using the FilterLine H Set Infant/Neonatal during MRI scanning could harm the
patient.
CAUTION:
During MRI scanning, the monitor must be placed outside the MRI suite. When the monitor
is used outside the MRI suite, EtCO
2
monitoring can be implemented by attaching the
FilterLine XL, to provide extended length.
CAUTION:
Use of a CO
2
sampling line with H in its name (indicating that it is for use in humidified
environments) during MRI scanning may cause interference.
The use of non-H sampling
lines is advised. For a list of H sampling lines, see
page
Non-invasive etCO
2
monitoring during magnetic resonance imaging (MRI) can be accomplished with the
monitor, a FilterLine XL, and an appropriate CO
2
sampling line.
T
O USE THE MONITOR DURING
MRI
SCANNING
:
1. Place the monitor outside the MRI suite. There must be a hole in the wall of the suite (approximately 10 cm.
diameter).
2. Connect the FilterLine XL to the monitor and guide the FilterLine XL through the hole in the wall of the
MRI suite.
3. Attach the FilterLine XL to the patient.
Note:
Due to the extended length of the FilterLine XL, there may be an increased delay time and thus a slower
response time.
To purchase the FilterLine XL, contact your local representative.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......