Installing the Battery Pack
Portable Bedside Capnograph/Pulse Oximeter
23
Ensure that the battery pack is fully charged before using the monitor without AC power. A fully charged
battery pack provides 2.5 hours of operation (without printer usage). When the monitor is connected to the AC
mains, the battery pack charges automatically. It takes approximately 12 hours to fully charge an empty battery
pack.
When you start using the monitor, verify that the battery icon at the bottom left of the monitor screen indicates
that the battery is full. Refer to
Testing the Battery and AC Connections
below for details.
If no battery pack is installed during monitor setup, the No Battery icon
will appear on the screen and the
advisory message
NO BATTERY INSTALLED
will appear.
WARNING:
It is recommended to always have a battery installed. If the battery is not installed, the unit
will operate properly on AC power, but if AC power is lost for any reason the monitor will
not work.
WARNING:
To replace the battery, first turn off the monitor and then unplug the unit from AC power.
Do not attempt to disconnect or connect a battery while the unit is turned on or connected to
AC power.
Note:
If the battery is not fully charged, the battery icon will indicate the charge
level of the battery.
Testing the Battery and AC Connections
The battery pack charge level and AC power connections should be confirmed before each use.
T
O TEST THE BATTERY
:
1. Press the ON/OFF
button to turn on the monitor.
2. Observe the battery icon level in the bottom left hand corner of the screen.
Figure 3 - Menu Bar with Battery Charge Level
3. If you have previously fully charged the battery, the battery icon should indicate that the battery is full.
Note:
As part of the monitor power-up, the battery charge level indicator will show full for about 15 seconds
after the monitor is turned on. The monitor will then update the battery charge level indicator to show the
true battery level.
Recharge the battery pack when the advisory message
BATTERY LOW
appears on the display screen. To
recharge the battery, make sure that the monitor is plugged into the AC mains. The orange AC power indicator
on the front panel of the monitor
will light up.
For normal operation, always check that the orange AC power indicator light is on during monitor use. This will
ensure the battery is charged during use and the monitor is prepared in case of a power outage or a patient
transfer. If a patient has to be transferred to another location, the unit can be unplugged and transferred with the
patient. Care should be taken to reconnect the monitor to the AC mains following the transfer.
Handling the Battery Pack
CAUTION:
Do not immerse the battery pack in water; it may malfunction.
CAUTION:
Recharge the battery pack only in the monitor to avoid possible heating, burning or rupture
of the battery pack.
Battery Charge
Level Indicator
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......