Compliance
Portable Bedside Capnograph/Pulse Oximeter
133
Table 32 - Guidance and Manufacturer's Declaration - Electromagnetic Emissions
Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF emissions EN 55011
Group 1
The Capnostream20P uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
Class A
The monitor is suitable for use in hospitals
EN 55011
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/
flicker emissions EN
61000-3-3
Clause 5
Table 33 - Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Immunity Test
IEC 60601
test level
Compliance
level
Electromagnetic Environment –
Guidance
Electrostatic discharge
(ESD)
EN 61000-4-2
± (2, 4, 6) kV contact
± (2, 4, 8) kV air
± (2, 4, 6) kV contact
± (2, 4, 8) kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Electrical fast
transient/burst
EN 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output lines
± 2 kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
EN 61000-4-5
± 1 kV differential
mode
± 2 kV common mode
± 1 kV differential
mode
± 2 kV common mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN 61000-4-11
<5 %
U
T
1
(>95 % dip in
U
T
)
for 0,5 cycle
<5 %
U
T
(>95 % dip in
U
T
) for
0,5 cycle
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
Capnostream20P requires continued
operation during power mains
interruptions, it is recommended that
the monitor be powered from an
uninterruptible power supply or a
battery.
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T
)
for 5 ses
<5 %
U
T
(>95 % dip in
U
T
)
for 5 sec
Power frequency
(50/60 Hz) magnetic
field
EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......