Troubleshooting
116
Portable Bedside Capnograph/Pulse Oximeter
CAUTION:
Do not spray or pour any liquid directly on the monitor, accessories or consumables.
CAUTION:
Do not use caustic or abrasive cleaners, or harsh solvents, including petroleum-based or
acetone solutions, to clean the device.
CAUTION:
Microstream
®
etCO
2
consumables are designed for single patient use and are not to be
reprocessed. Do not attempt to clean, disinfect or blow out the FilterLine as the monitor can
be damaged.
CAUTION:
Do not expose SpO
2
sensor connector pins to cleaning solution as this may damage sensor.
CAUTION:
If a 1:10 bleach dilution (0.5% to 1% Sodium Hypochlorite solution) in the form of wipes is
used to disinfect external surfaces of the monitor, exposure of the connectors and the
display to the bleach solution should be avoided. Repeated cleaning with bleach over time
may cause discoloration and residue on the surfaces.
Troubleshooting
This section lists potential problems you may experience while using the monitor and suggestions for resolving
them. If you are unable to correct the problem, contact qualified service personnel or your local representative.
Electrical
Problem
Cause
Action
Monitor does not turn
on.
Internal battery is totally discharged
and power cable improperly
attached or disconnected, or cable
has faulty electrical connection.
Check power cable connection.
AC wall outlet has no power and
internal battery is not charged.
Check connections and correct problem.
Blown fuses.
Replace the fuses. Contact your
representative to determine the reason for the
electrical problem.
AC mains power and
monitor on indicator
lights are on, but unit
will not operate on
battery power when the
AC mains power cable
is disconnected.
Battery pack not plugged in to
monitor.
Open the battery housing and check that the
battery pack cable is firmly connected to the
battery socket. (See
Monitor is plugged in,
but does not appear to
charge the battery.
AC power not getting to the monitor.
Check if the orange power indicator light is on.
If not, check that the AC power cord is
properly plugged in to a live AC mains socket.
Battery is not fully
charged after being
plugged in for 24 hours.
Battery pack has not recharged
completely.
Unplug the monitor from AC power for 3-4
hours, and then plug in again. The battery
pack will now recharge completely when
connected to AC power.
If the battery pack still does not recharge
completely, replace the battery pack.
CO
2
Problems
Problem
Cause
Action
NO BREATH message
appears constantly and
red alarm indicator
flashes.
Physiological cause.
Check patient.
Clogged or blocked FilterLine.
Check FilterLine and replace if blocked.
FilterLine caught in something or
tube is kinked.
Check the FilterLine from the monitor all the
way to the patient to see if line is kinked,
twisted closed or caught in bed or equipment.
FilterLine connected
FilterLine not plugged in properly.
Check that the FilterLine plug is screwed into
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......