Parameter Standby Mode
76
Portable Bedside Capnograph/Pulse Oximeter
Message
Description
USB DEVICE FAILED
A USB device connected to the monitor has failed.
REPORT TRANSFER COMPLETE
Data communication is complete.
CO
2
MONITORING HAS BEEN
OFF FOR HH:MM
Displays the hours and minutes the pump has been turned off
during PUMP OFF mode.
DEMO MODE - PRERECORDED
DATA
Displayed during demo mode when no other message is
displayed.
SPO2 LOW ALARM LIMIT: XX
Displayed if the SpO
2
Low alarm limit is set below 85%.
REMOTE SYSTEM CONNECTED*
Capnostream is connected to a remote system. This message
will only be present if enabled by the host computer, and may
have different wording if so programmed by the host computer.
REMOTE SYSTEM
DISCONNECTED*
Capnostream is no longer connected to a remote system. This
message will only be present if enabled by the host computer,
and may have different wording if so programmed by the host
computer.
INCOMPATIBLE SOFTWARE
VERSION
Displayed during transfer of institutional defaults
NO FILE FOUND
Displayed during transfer of institutional defaults
A/hr REQUIRES 1 HOUR OF
DATA (if available)
A/hr does not appear, since one hour of data is required to
compute this value, and one full hour of data is not yet available
ODI REQUIRES 1 HOUR OF DATA
(if available)
ODI does not appear, since one hour of data is required to
compute this value, and one full hour of data is not yet available
CO
2
STANDBY
CO
2
standby has been activated.
SPO
2
STANDBY
SpO
2
standby has been activated.
NO BATTERY INSTALLED
No battery is installed in the device
* When used with a remote system, this message may be displayed with a different wording if so programmed by
the host computer. The host computer may also initiate a message which would appear upon the stopping of
communications.
Parameter Standby Mode
There is an option of placing the monitor in a separate parameter standby mode for capnography and for pulse
oximetry. Once in this mode, the user can activate Parameter Standby as required. The purpose of this option is
to enable a situation in which the device will alarm when a sampling line/sensor is disconnected from the
device, but permit the user to turn off this option at will.
The options are as follows:
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......