Capnostream®20p: Operational Check Sheet
40
Portable Bedside Capnograph/Pulse Oximeter
Capnostream
®
20
p
: Operational Check Sheet
To get the Capnostream monitor up and running quickly and smoothly, follow the list of instructions below:
1.
Unpack the monitor
•
Remove the Capnostream monitor and the accessories from the box.
•
Check that the items on the enclosed packing list are included.
•
Check that paper is in printer.
2.
Install the battery pack
•
for installation instructions.
3.
Turn on the monitor
•
Plug the power cord into the mains plug in the rear of the monitor. Clip the power cord strain relief
around the cable and tighten to ensure that the power cord does not accidentally disconnect from the
monitor.
•
Plug the power cord into the mains AC supply.
•
The power indicator orange light at the front of the monitor will turn on.
•
Press the Power ON/OFF button
on the front panel to turn on the monitor. The power-on
indicator green light will light up, showing that the monitor is turned on.
•
Note that both the power indicator orange light and the power-on indicator green light should both
be on during operation of the monitor.
4.
Change the date, time or language
•
Use the control knob to select
SYSTEM
in the Menu Bar
•
Change each setting on that screen, if desired.
5.
Set the Patient Type and Mode
•
Use the control knob to select the
PATIENT TYPE
function on the menu bar of the Home screen
•
Select
ADULT
,
PEDIATRIC
(by age), or
INFANT/NEONATAL
as appropriate.
6.
Connect a FilterLine
•
Slide open the FilterLine input connector shutter and connect the appropriate FilterLine.
•
Connect the FilterLine to the patient as described in the Directions for Use supplied with the
FilterLine. The sampling line connecter should be screwed clockwise into the monitor CO
2
port
until it can no longer be turned, to ensure that it is connected securely to the monitor. This will
assure that there is no leak of gases during measurement at the connection point and that
measurement accuracy is not compromised.
7.
Connect a SpO
2
Sensor
•
Connect the SpO
2
extension cord firmly to the monitor SpO
2
sensor port, and then connect the
appropriate SpO
2
sensor to the extension cord.
•
Connect the SpO
2
sensor to the patient as described in its Directions for Use.
8.
Once either or both sensors are connected to the monitor, it is ready for operation.
9.
Check Alarm Limits
•
Check the alarm limits default and make any permanent changes in the institutional defaults.
•
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......