Alarm Delay
48
Portable Bedside Capnograph/Pulse Oximeter
CAUTION:
The Audio off option (which will create permanent alarm silence) should be set in the
Institutional Defaults only in a situation in which the caregiver is also monitoring the
patient by other means, to avoid the chance of missed alarms.
Alarm Delay
An option for Alarm Delay (for a number of alarms) is available. For more information, see
page
Use of Scavenging System
When the patient is being sedated with a gaseous anesthetic, a scavenging system can be attached to the
monitor. The gas output connection is a barbed style connector intended for 3/32 inch ID tubing. Using any
appropriate tubing, connect the scavenging system to the gas outlet, located between the FilterLine and SpO
2
connections, as shown below in
Disposal of sampled gases should be carried out according to according to standard operating procedures or
local regulations for the disposal of gases.
Figure 19 - Scavenger System Connection Point
Turning the Pump Off for Suction or Lavage
Use the Pump Off mode whenever performing suction or lavage. During Pump Off mode, pump activity is
suspended to protect the monitor from drawing in liquids which could cause a malfunction.
In the Pump Off mode, the CO
2
module pump is switched
OFF
for a preset time to prevent liquids from
entering the monitor.
WARNING:
If at any time the device displays the
FILTERLINE BLOCKAGE
message, replace the
FilterLine.
T
O CHANGE THE PUMP MODE
:
1. Before performing lavage or suction, press the Pump Off button
located on the front of the monitor.
The CO
2
module pump turns off, the countdown timer begins and the Pump Off screen is displayed. The
countdown timer is shown in the CO
2
waveform area.
Scavenger system Connection Point
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......