Changing the Alarm and Pulse Volumes
Portable Bedside Capnograph/Pulse Oximeter
47
Figure 16 - Menu Bar
2. Click the control knob once to select the alarm volume control.
3. Turn the control knob to raise or lower the volume. The selected alarm volume level will be audible as you
turn the knob. Click the knob twice to set the new volume level.
Figure 17 - Alarm Volume Selection
Note:
The alarm volume cannot be set to zero using the alarm volume control. The audible alarm can only be
disabled in the Institutional Default settings.
Pulse Tone Volume
The monitor can be set to sound an audible beep for each pulse beat. The monitor is shipped from the factory
with the pulse tone turned off.
The pulse tone can also be turned off using the SpO
2
menu. By default, the pulse tone is turned off.
Setting the Pulse Tone Volume
To make the pulse tone louder or softer, use the control knob to select the speaker icon and click twice to select
the Pulse Tone volume setting.
Figure 18 - Pulse Tone Volume Selection
The pulse tone volume can be set to zero.
Turning the Pulse Tone On/Off
T
O TURN THE PULSE TONE VOLUME ON
:
1. Use the control knob to select the SpO
2
display area. Click the control knob to go to the SpO
2
Setup screen.
2. Turn the control knob to highlight the
PULSE TONE
setting, click the knob to select the option, turn the
knob once to change the setting to ON, and push the knob again to set the option.
3. A beep now sounds once for each pulse beat. To turn the pulse tone off, repeat the procedure and select
PULSE TONE
off.
Alarm Volume Default Options
The institutional default alarm volume can be set to one of the three choices described below, in the
Institutional Defaults>Monitor
screen. These choices determine the audio alarm volume when the device is
on page
to learn more about Institutional Defaults.
Table 9 - Audio Alarm Volume
Audio Alarm Volume Option
Description
Maximum
The audio alarm volume will revert to the maximum volume when
the device is turned on, regardless of previous settings.
Last Setting
The audio alarm volume will remain at the level last set on the
device, even after it is turned off and then turned on again.
Audio Off
The audio alarm volume is off.
Speaker
Volume
icon
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......