Data Transfer via the USB Data Port
104
Portable Bedside Capnograph/Pulse Oximeter
Report Name
Description
Fields Included
Real Time Continuous
CO
2
Waveform
Tab delimited (.txt) file with data entries
every 50 milliseconds.
DATE, TIME
Patient Reading every 50 milliseconds (for
creating CO
2
waveform): CO
2
*
Real Time Continuous
Tabular
Tab delimited (.txt) file, similar to the
Tabular Trend report, but transmitted line
by line in real time.
DATE, TIME
Patient Readings: EtCO
2
, FiCO
2
, RR, SpO
2
, PR,
IPI, A/hr, ODI.
Patient Urgent and Caution Alarm Occurrences
Equipment Advisory Message Occurrences
Events: EVENT 1, EVENT 2, EVENT 3
Real Time Continuous
Tabular with real-time
continuous waveform
(shortened name is
FULL CONTINUOUS
TRANSFER)
Tab delimited (.txt) file, similar to the real
time continuous tabular report, transmitted
line by line in real time, but with data entries
every 50 milliseconds (20 times a second).
CO
2
data which can be used to create a
real-time continuous waveform is shown,
with data entries every 50 milliseconds.
Tabular data is recorded at data points
based on the resolution set for trend
storage (5, 10 or 30 seconds), so data
between these points is repeated on
additional rows.
DATE, TIME
Patient Readings (at trend storage resolution):
EtCO
2
, FiCO
2
, RR, SpO
2
, PR, IPI, A/hr, ODI
Patient Urgent and Caution Alarm Occurrences
Equipment Advisory Message Occurrences
Events: EVENT 1, EVENT 2, EVENT 3
Patient reading every 50 milliseconds (for
creating CO
2
waveform): CO
2
wave
Full Binary Continuous
Transfer
See the document Capnostream Data Transfer Protocols
Full Binary Trend
Transfer
See the document Capnostream Data Transfer Protocols
*
CO
2
in mm/Hg (millimeters of mercury)
Note that in the .txt files, the first six lines of data are as follows:
Line 1 – The name of the report type.
Line 2 – Blank, or Patient ID if the report is a case report
for information on patient types)
Line 4 – Blank
Line 5 – Column headings
Line 6 – Column headings second line
Capnostream recognizes flash memory drive devices manufactured by SanDisk, Lexar, and PNY Technologies.
It will not recognize flash memory drives that have advanced capabilities which require the installation of
additional drivers, such as the SanDisk U3. The capacity limit for a USB drive that can be used with
Capnostream is 8 GB. A typical flash memory device is shown below.
Figure 40 - Typical Flash Memory Device
T
O
R
ECORD
C
APNOSTREAM
D
ATA ON A
USB
D
EVICE
:
1. Insert a USB Flash Memory drive into the USB port on the back side of the Capnostream.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......