Warnings
64
Portable Bedside Capnograph/Pulse Oximeter
monitored during sedation or pain management) may require a lower IPI Low Alert threshold to reflect
their impaired respiratory capacity.
The IPI is available for all three groups of pediatric patients (1-3 years, 3-6 years, and 6-12 years), and for adult
patients. It is not available for Neonatal/Infant patients (patients up to the age of one year), and thus will not
appear on screens for Neonatal/Infant patients.
Warnings
WARNING
:
Ensure that the patient type is correctly selected before beginning monitoring of a patient.
Choosing an incorrect patient type could produce incorrect patient IPI data.
WARNING:
When an IPI Low Alert is triggered for a patient, medical staff should review the patient's
status to determine if a change in medical care is required.
IPI Display
IPI appears on the Home screen and is available in all Capnostream functions as a default option, along with
other patient parameters such as etCO
2
or SpO
2
. On the Home screen, both numerics and a trend graph are
provided for IPI.
The IPI option may be disabled from the Institutional Defaults screen; see below for more information. In the
Neonatal measurement mode, the IPI option is automatically disabled.
IPI Options
T
O
C
HANGE
IPI
O
PTION
S
ETTINGS
:
1. On the Home screen, click the
SYSTEM
button in the menu bar at the bottom of the screen.
2. The System Setup screen will appear. Move the control knob to the parameter that you wish to change and
click to select that parameter. Move the control knob to select the desired setting and click to select the
setting. A list of the IPI settings that can be changed appears in
Table 13 - Adjustable IPI Options
3. Move the control knob to the
HOME
button and click to select. The screen will also revert to the Home
screen after a few seconds if no additional actions are taken on that screen.
4. If the IPI Alert is disabled, it will not appear in the Home screen or in any other screen. It will be listed but
grayed out in the Alarm Limits screen.
5. The changes in the parameters will remain in effect until the device is turned off.
Table 13 - Adjustable IPI Options
Parameter
Choices
Factory Default
Home IPI Display
1 hour, 2 hour, 4 hour
1 hour
IPI Alert
Enabled/Disabled
Disabled
To change IPI Display and Alert options on a more permanent basis, use the Institutional Defaults option,
described in
on page
. The Home IPI Display settings can be changed in the
Institutional Defaults: Trend screen and the IPI Low Alert (enabled or disabled) can be changed in the
Institutional Defaults: Monitor screen. IPI Low Alert Alarm Limits can be changed in the Institutional
Defaults:Alarm Limits:Set Adult/Pediatric screen.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......