Connecting a FilterLine
54
Portable Bedside Capnograph/Pulse Oximeter
For further information, please contact your local representative.
Select the appropriate FilterLine and connect it to the monitor before attaching it to the patient’s airway. Be sure
to follow Microstream
®
etCO
2
Consumables’
Directions for Use
for proper connection.
Microstream
®
EtCO
2
Consumables
A list of Microstream
®
etCO
2
on page
Note:
Smart products provide oral and nasal sampling.
H products are for long term use.
Note:
The generic term FilterLine, used in this manual, is interchangeable with any of the Microstream
®
etCO
2
consumables.
Note:
The listed products are also available in versions designed for other patient sizes.
This listing describes the main products available. For more information about Oridion FilterLines or additional
sizing and packaging options for these products, see http://www.covidien.com/rms/brands/microstream.
Connecting a FilterLine
Before monitoring a patient with capnography, the appropriate FilterLine must be connected to the monitor and
to the patient.
T
O MAKE THE CONNECTIONS
:
1.
Slide open the FilterLine input connector shutter and connect the appropriate FilterLine. Screw the
FilterLine connector into the monitor clockwise until it can no longer be turned.
2.
Connect the FilterLine to the patient as described in the
Directions for Use
supplied with the
FilterLine.
When the FilterLine is connected, the monitor will immediately begin to search for breaths, but it will not
indicate a No Breath condition before any valid breaths have occurred.
Note:
The Capnostream monitor’s CO
2
function will enter standby mode automatically in cases when the
Capnostream monitor is left on for 30 minutes or longer without a FilterLine attached. In this case, CO
2
monitoring on the monitor will go into automatic standby mode and then will re-initialize once a FilterLine
is attached. This initialization process takes typically 30 seconds; it may take up to 180 seconds.
CO
2
Data Displayed by the Capnostream Monitor
The Capnostream monitor Home screen displays real time CO
2
data. The displayed data includes:
•
Real time etCO
2
and FiCO
2
values along with selected unit (see
on page
for details regarding the available units)
•
Real-time FiCO
2
values along with selected unit (see
on page
for
details regarding the available units)
•
Respiration rate (RR) in breaths per minute
•
CO
2
Waveform
Figure 22 - CO
2
Data on the Capnostream Monitor
CO
2
Waveform
EtCO
2
, FiCO
2
and Respiration
Rate values
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......