Data Transfer via the USB Data Port
Portable Bedside Capnograph/Pulse Oximeter
105
2. When the Flash Memory drive is detected the USB icon will appear in the top right-hand corner of the
display beside the alarm symbol. Depending on the type of drive, this may take up to 40 seconds.
Figure 41 - USB Icon
Note:
The USB port on the Capnostream monitor is for use with a Flash Memory device only. It is not a full
service USB port. Do not attempt to connect the monitor to a computer via the USB port.
Note:
The USB Flash Memory drive must be inserted with care in the USB port, without using excessive force.
If the Flash Memory drive cannot be inserted easily into the port, do not use the drive.
3. After the USB icon appears, the monitor is ready to begin outputting data to the USB flash memory device.
4. From the Home screen select the
SYSTEM
button on the menu bar to open the System screen, and then
select
DATA OUTPUT
.
5. Use the control knob to select the desired report from the
DATA OUTPUT
table as shown below. Please
note that the Tabular Case and Tabular Trend options are only available while a case is active. If the current
case is closed, the case and trend memory are deleted, and this information will no longer be available.
Table 25 - Select Data Output Type
TABULAR CASE
TABULAR TREND
REALTIME CONTINUOUS CO
2
WAVEFORM
REALTIME CONTINUOUS TABULAR
FULL CONTINUOUS TRANSFER
FULL BINARY CONTINUOUS TRANSFER
FULL BINARY TREND TRANSFER
6. An asterisk will appear to the left of the selected report name. If no case is active, the text NO CASES will
appear to the right of the Tabular Case option when that option is selected.
7. Turn the control knob to select
START USB
on the Menu Bar and click to begin data transfer. Data output
can be aborted by clicking again to select
STOP USB
.
CAUTION:
If the Flash Memory disk drive is removed from the Capnostream when data transfer is in
progress, the data may not be readable. Before removing the Flash Memory drive, data
transfer should be completed or stopped by selecting
STOP USB
on the Menu Bar.
Note:
If the Capnostream does not detect the Flash Memory drive, remove and re-insert the Flash Memory
drive. If the Flash Memory drive is still not detected, check that drive being used is from a supported
manufacturer.
Note:
If free disk space on the Flash Memory drive is less than 100 kb, writing to the USB disk drive is not
allowed. Under this condition, if data transfer is already in progress, it will be aborted. Any new data
transfer CANNOT be initiated under the low disk space condition.
Note:
Please note additional details regarding trend reports in
Important Notes Regarding Trend Reports
The maximum amount of data that can be transferred in a single file is 65,536 rows (this corresponds to the
maximum sheet size for an Excel file for Excel 2003 and lower). If the data exceeds 65,536 rows, a new file is
automatically opened and the data continues to transfer into the new file. In this situation, the new file name is
indexed as described below in
Table 26 - File Naming Conventions
The following are estimates of the approximate sizes of the files that can be expected to be generated. For
patient cases where events and alarms are recorded extensively, the file sizes will be larger.
Tabular Case: 1 hour @30s resolution: 21kB
USB port icon
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......