Portable Bedside Capnograph/Pulse Oximeter
53
Do not touch this field
- it is invisible and does not appear in the
final document
Chapter 5
Capnography with the Capnostream Monitor
CO2 Data Displayed by the Capnostream Monitor
Monitoring CO2 during MRI Scanning
Microstream
®
EtCO
2
Consumables
CAUTION:
Before use, carefully read the Microstream
®
etCO
2
consumables
Directions for Use
.
CAUTION:
Only use Microstream
®
etCO
2
consumables to ensure the monitor functions properly.
CAUTION:
Microstream
®
etCO
2
consumables are designed for single patient use, and are not to be
reprocessed. Do not attempt to clean, disinfect or blow out the FilterLine as the monitor can
be damaged.
CAUTION:
Dispose of Microstream
®
etCO
2
consumables according to standard operating procedures or
local regulations for the disposal of contaminated medical waste.
WARNING:
Loose or damaged connections may compromise ventilation or cause an inaccurate
measurement of respiratory gases. Securely connect all components and check connections
for leaks according to standard clinical procedures.
WARNING:
If too much moisture enters the FilterLine (i.e., from ambient humidity or breathing of
unusually humid air), the message
Clearing FilterLine
will appear in the Capnostream
message area. If the FilterLine cannot be cleared, the message
FilterLine Blockage
will
appear in the CO
2
waveform display section on the Home screen and in the Capnostream
message area. (If there is no waveform display, the message will appear only in the message
area.) Replace the FilterLine once the
FilterLine Blockage
message appears.
WARNING:
The FilterLine may ignite in the presence of O
2
when directly exposed to laser, ESU
devices, or high heat. When performing head and neck procedures involving laser,
electrosurgical devices or high heat, use with caution to prevent combustion of the
FilterLine or surrounding surgical drapes.
Note:
When connecting a sampling line to the monitor, screw the sampling line connecter clockwise into the
monitor CO
2
port until it can no longer be turned, to ensure that it is connected securely to the monitor.
This will assure that there is no leak of gases during measurement at the connection point and that
measurement accuracy is not compromised.
Note:
Following connection of the CO
2
sampling line, check that CO
2
values appear on the monitor display.
Basic Principles
When choosing Microstream
®
etCO
2
consumables, the following should be considered:
•
Whether the patient is intubated or non-intubated
•
Whether the patient is on mechanical ventilation
•
Duration of use
•
Patient’s size and weight
•
Whether the patient is breathing through his nose, his mouth, or alternating between oral and nasal
breathing
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......