Clearing Trend Memory
90
Portable Bedside Capnograph/Pulse Oximeter
Clearing Trend Memory
It is recommended to erase trend memory when the monitor is switched to a new patient, in order to avoid
mistaking the earlier data for the present patient's data. If you are working with cases, and the current case is
ended, the trend memory is automatically cleared.
T
O
E
RASE
T
REND
M
EMORY
:
1. To erase the trend data from the monitor memory, use the control knob to go to the Home screen and select
SYSTEM
on the menu.
2. On the System screen, select
CLEAR TREND
. The word
CONFIRM?
will appear just above the Menu Bar.
3. If you are certain that you want to erase the trend memory, click the control knob. If you do not want to
erase the trend memory, turn the knob to the left or right to cancel.
4. When the machine is turned on, a message appears, suggesting that you erase the trend memory in order to
start a new patient without information from previous patients in the trend memory. This screen appears in
Figure 34 - Trend Memory Message
, below. Click
YES
to erase the trend memory. If you intend to continue
measurement of the same patient as previously, you may want to retain the trend memory. In that case, click
NO
. If you record patient data as part of a case, trend memory will always be erased when you close the
case.
Figure 34 - Trend Memory Message
Configuring Trends
To change the parameters for trend displays, go to the Home screen and select
SYSTEM
to see the System
Setup screen. The table below shows the options as they appear on the System Setup screen.
Table 22 - Monitor Parameters
DATE
MAY 25, 2011
TIME
11:27:32 AM
LANGUAGE
ENGLISH
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......