Parameter Standby Mode
Portable Bedside Capnograph/Pulse Oximeter
77
In Standard Mode (when Parameter Standby Mode is not enabled), removing a FilterLine or SpO
2
sensor/extension cable from the device will cause a message to appear on the screen (
FILTERLINE
DISCONNECTED
or
SpO
2
SENSOR DISCONNECTED
, as relevant) but no alarms will sound. Removing
the SpO2 sensor from patient will set off an audible alarm and an on-screen message. By default, Parameter
Standby mode is disabled.
When Parameter Standby Mode is enabled, after a FilterLine has been connected and then removed from the
device, a medium priority alarm
FILTERLINE DISCONNECTED
will sound. Likewise, after a pulse oximetry
sensor/extension cable has been connected and then disconnected from the device, a medium priority alarm
SPO
2
SENSOR DISCONNECTED
will sound. The purpose of this alarm is to prevent unauthorized
disconnection of a FilterLine/SpO
2
sensor from the device, perhaps by patients or patients’ visitors.
When Parameter Standby Mode is enabled, removing the SpO
2
sensor from patient will set off an audible alarm
and an on-screen message, as it does in Standard Mode.
Parameter Standby mode can be enabled by entering
SYSTEM>SERVICE>
Enter Service password
>INST
DEFAULTS>MONITOR>
Turn
PARAMETER STANDBY MODE
to
ENABLED
. Return to the Home screen
by clicking
BACK>BACK>HOME
.
However, there may be situations in which the clinician wants to use the Parameter Standby Mode but does not
want these alarms to sound, for either a disconnected FilterLine or an SpO
2
sensor, because they have
authorized the disconnection. For these situations, the monitor provides an option to put one or both parameters
on standby while these activities take place.
Once Parameter Standby mode is enabled under institutional defaults, it can be made active as follows:
1. Remove the FilterLine and/or SpO
2
sensor from the device or remove the SpO
2
sensor from the patient.
Note:
The capnography parameter standby mode can be made active only if the FilterLine Disconnected alarm
is currently alarming, or if a FilterLine has not yet been connected to the device, since the time that it
was turned on. Likewise, the pulse oximeter parameter standby mode can be accessed only if the SpO
2
Sensor Disconnected alarm or the SpO
2
Sensor Not on Patient alarm is currently alarming, or if an SpO
2
sensor has not yet been connected to the device or sensed a pulse, since the time it was turned on.
Note:
The capnography parameter standby mode cannot be activated if the monitor is currently monitoring a
patient with capnography. Likewise, the pulse oximeter parameter standby mode cannot be activated if
the monitor is currently monitoring a patient with pulse oximetry.
2. To activate Parameter Standby Mode, press the alarm silence button at the bottom of the front panel (
)
for 2 seconds or longer. A new message
CO
2
STANDBY
and/or
SPO
2
STANDBY
(as required) will now
appear in the waveform area (if the waveform area is available) and in the message area. (This is in addition
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......