Portable Bedside Capnograph/Pulse Oximeter
119
Do not touch this field
- it is invisible and does not appear in the
final document
Appendix 1
Institutional Settings
Changing Institutional Defaults
Uploading or Downloading Institutional Defaults
Institutional Defaults
Capnostream is shipped from the factory with all changeable settings configured according to the tables in the
section
on page
. These are called the Factory Default Settings. If the specific
environment of use indicates that other settings are preferable or required, or Institutional policy requires
different values than the Factory Defaults, then the default settings can be changed so that they are in effect
every time the monitor is turned on. This is more reliable than expecting staff members to change the settings
before each use.
Default settings can be changed manually by an authorized technician / biomedical engineer to produce
institutional defaults. Default settings are set from the Institutional Defaults screen, which is reached from the
Service screen. The Service Screen is password protected. The procedure is described in
Changing Institutional Defaults
From the Home screen use the control knob to select the
SYSTEM
button, and then select the
SERVICE
button on the menu to open the Service screen. The enter password box appears near the top of the screen.
ENTER PASSWORD
As described in
, use the control knob to select
ENTER PASSWORD
. Using the
control knob, enter the password and click the control knob again after the last letter. The password appears in
on page
. The service password page may be removed from the manual for
safekeeping.
Use the control knob to select
INST DEFAULTS
. You have the option of changing the default settings for
ALARM LIMITS, TRENDS, MONITOR, CO
2
and SpO
2
.
WARNING
:
Changing the settings might adversely affect the monitoring of patients. Changes to the
Institutional settings must only be made by authorized personnel.
Note:
Patients will not be monitored by the device while it is in service mode, even if the monitor is connected
to a patient while in service mode. (The monitor is in service mode once the service password is
inputted.) Therefore, you may want to remove the FilterLine from the patient or disconnect the FilterLine
from the monitor while putting the monitor into service mode. No data will be recorded while the monitor
is in service mode, and thus trying to monitor while in service mode will lead to missing data issues.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......