Changing Monitor Settings
Portable Bedside Capnograph/Pulse Oximeter
123
Alarm Delay
A number of alarms can be delayed by a choice of 10, 15, 20, or 30 seconds, or not at all (Alarm Delay
disabled). This option is available for all patient types.
The Alarm Delay can be set from the Institutional Defaults screen, which can be reached by
SYSTEM>SERVICE>INPUT SERVICE PASSWORD
(see
>INST DEFAULTS>ALARM LIMITS>SET ALARM DELAY
.
Table 30 - Factory Default and Optional Alarm Delay Settings
Parameter
Choices
Factory Default
EtCO
2
High Alarm Delay
10, 15, 20, or 30 seconds or Disabled
DISABLED
RR High Alarm Delay
10, 15, 20, or 30 seconds or Disabled
DISABLED
SpO
2
Low Alarm Delay
10, 15, 20, or 30 seconds or Disabled
DISABLED
PR Low Alarm Delay
10, 15, 20, or 30 seconds or Disabled
DISABLED
PR High Alarm Delay
10, 15, 20, or 30 seconds or Disabled
DISABLED
Trend Settings
Institutional Defaults can be set to change the Trend stored in the monitor, and how it is displayed on the screen.
For the specific trend settings for SpO
2
and CO
2
, see the individual settings described below for the SpO
2
and
CO
2
parameters.
Table 31 - Factory Default and Optional Trend Settings
Parameter
Choices
Factory Default
Trend Recording
Resolution
5 seconds at 12 hours
10 seconds at 24 hours
30 seconds at 72 hours
30 seconds at 72 hours
Tabular Increment
Display Default
MINIMUM, 1.5, 3, 15 or 60 minutes
1.5 Minutes
Graphical View Default
2, 4 or 12 hours
4 hours
The Home IPI Display Default (see
) also appears in the Trend Settings defaults screen.
The default Trend Recording Resolution determines how many hours of patient information can be recorded.
The Tabular Increment Display enables the default to be set to the values in the table.
The Graphical View Default can be changed so that the Trend window will show a different time period of data.
Note:
Changing the Trend resolution will clear the Trend memory, erasing any patient data that was in the
monitor.
To make changes to the Trend Defaults, click
SERVICE>INST DEFAULTS>TREND
on the menu bar. Use
the control knob to navigate to the parameters described above and click to view options. Chose an option using
the control knob and click again to record that option as the default.
Changing Parameters Order on the Trend Display
The Graphical and Tabular screens show five different patient parameters. The order of the trend parameters can
be changed in the Institutional defaults screen, in order to show the more relevant parameters on the first trend
display screen rather than the second screen. The process is as follows: Click
SYSTEM>SERVICE>
Enter
)>
INST DEFAULTS>TREND>CONFIG.
DISPLAY
to open the Institutional Defaults: Trend: Display Configuration screen.
For each position that you wish to change, move the cursor to that line with the control knob, click the control
knob, and then choose, from the Parameters Available column, the parameters which you wish to show in that
line. For example, in Parameter Position 1, the default is EtCO
2
. To change the trend display so that IPI appears
as the first parameter on the trend screen, move the cursor to Parameters Selected: EtCO
2
, click the control
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......