Alarm Silence
78
Portable Bedside Capnograph/Pulse Oximeter
to the
FILTERLINE DISCONNECTED
or
SPO
2
SENSOR DISCONNECTED
or
SpO
2
SENSOR NOT
ON PATIENT
alarms, as relevant, which will continue to appear in the waveform and message areas.)
Note:
A short press on the Alarm Silence button will still turn off the audible alarms temporarily (for 2 minutes),
as previously.
3. Once parameter standby is activated, the following will take place: In this mode, CO
2
and/or SpO
2
alarms do
not exist, so neither audible alarms, the flashing yellow LED on the Capnostream front panel, storage of
these alarms in any data storage method, nor transfer of these alarms to central stations, nurse call
indications, etc., will take place.
4. To access both parameter standby modes at the same time, enter them one at a time, or press the Alarm
Silence button (
) while the FilterLine is disconnected and the SpO
2
sensor is not connected to a patient
or device, to enter both at the same time.
To exit the capnography parameter standby mode, reattach a FilterLine. To exit the pulse oximetry parameter
standby mode, reattach the SpO
2
sensor to the patient and device. (Thus, if the user is unaware that the monitor
is in the Parameter Standby mode, simply reattaching a FilterLine to the monitor or SpO
2
sensor to the patient
will cause the monitor to exit the Parameter Standby mode and monitoring to resume.)
Table 19 - Message and Alarm Status during Different Parameter Standby Situations
Feature
Status when
Parameter Standby
mode disabled
Status when
Parameter Standby
mode enabled but not
activated
Status when Parameter
Standby mode enabled and
Parameter Standby activated
FilterLine disconnected (from
device) on-screen message /
SpO
2
sensor disconnected
(from device) on-screen
message
Yes
Yes
Yes
FilterLine disconnected (from
device) alarm / SpO
2
sensor
disconnected (from device)
alarm
No
Yes
No
SpO
2
Sensor not on Patient
Medium Priority audible alarm
Yes
Yes
No
SpO
2
Sensor not on Patient
on-screen message
Yes
Yes
Yes
CO
2
Standby message / SpO
2
Standby message
No
No
Yes
High Priority (patient) alarms
Yes
Yes
No(for the parameter in standby)
Flashing red and yellow LEDs
on the Capnostream front
panel during high Priority
(patient) alarms
Yes
Yes
No, for the parameter in standby
(since high priority [patient]
alarms related to the parameter
in standby do not exist in this
case)
Storage or transfer to remote
stations of high priority
(patient) alarms
Yes
Yes
No, for the parameter in standby
(since high priority [patient]
alarms related to the parameter
in standby do not exist in this
case)
Alarm Silence
To temporarily silence/disable an alarm, press the alarm silence button:
.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......