Types of Alarms
74
Portable Bedside Capnograph/Pulse Oximeter
Message
Description
Corrective Action
SPO
2
LOW ALARM
The SpO
2
is below the lower
alarm limit
Patient requires immediate
medical attention.
PULSE RATE HIGH ALARM
The pulse rate is above the
upper alarm limit
Patient requires immediate
medical attention.
PULSE RATE LOW ALARM
The pulse rate is below the
lower alarm limit
Patient requires immediate
medical attention.
FiCO
2
HIGH ALARM
The FiCO
2
is above the upper
alarm limit
Patient requires immediate
medical attention.
IPI LOW ALARM*
The IPI is below the lower alarm
limit.
Patient requires immediate
medical attention.
SYSTEM RESET, CHECK
SETTINGS
The system has reset and will
revert to institutional default
settings.
Reset your alarm limits and
other settings to match the
requirements of the patient
being monitored.
*
The IPI LOW ALARM is an alert which is intended to indicate a change in patient status to the physician.
When this alert appears, levels of other patient parameters should be evaluated.
Medium Priority Alarms
Table 16 - Medium Priority Alarms
Message
Description
Corrective Action
CO
2
ERROR
Failure has occurred which
prohibits the operation of the CO
2
function.
Contact Oridion authorized
personnel.
SPO
2
ERROR
Failure has occurred which
prohibits the operation of the
SpO
2
function.
Contact Oridion authorized
personnel.
PULSE NOT FOUND
No detectable pulse.
Patient requires immediate medical
attention. Reposition sensor on
patient.
FILTERLINE
DISCONNECTED
No FilterLine is connected to the
device.
Will occur only when Parameter
Standby is enabled.
Insert FilterLine into CO
2
port on
device.
SPO
2
SENSOR
DISCONNECTED
No SpO
2
sensor is connected to
the device.
Will occur only when Parameter
Standby is enabled.
Insert SpO
2
sensor into SpO
2
port on
device.
SPO
2
SENSOR NOT ON
PATIENT
Sensor is off patient.
Place sensor properly on patient.
REPLACE SpO
2
CABLE
The SpO
2
cable is defective.
Replace the SpO
2
cable.
FILTERLINE BLOCKAGE
FilterLine is kinked or clogged.
Disconnect and reconnect the
FilterLine. Check the airway adapter
and if necessary, replace the
FilterLine.
BATTERY LOW
Battery charge level is low and
monitor will shut down soon.
Connect monitor to AC power.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......