Chapter 11
Reports
Apnea and O2 Desaturation Report
When the user clicks on the
REPORTS
button on the Home screen, the Apnea and O
2
Desat (Desaturation)
Report will appear on the screen. See
Apnea and O2 Desaturation Report,
below for more information. To print
other patient reports, click on the
PRINT REPORT
button on the Apnea and O
2
Desat Report screen. For more
If A/hr and ODI display is not available or disabled (either in Institutional Defaults or in the System Setup
screen), or if the Patient Type is set to Infant/Neonatal or one of the Pediatric patient types, the
REPORTS
button will not appear. Instead, clicking the
button on the Home screen will open the Print screen. For
Apnea and O
2
Desaturation Report
The on-screen and printed Apnea and Desat reports show a comprehensive picture of the number of apnea and
O
2
desaturation events over a set time period of 2, 4, 8, or 12 hours. They show the average A/hr and ODI score
for the whole period as well as listing the number of events during each part of the set time period, with the
apnea events listed by the number of seconds each event lasted. The sum of events in each row over the entire
time period (2, 4, 8, or 12 hours, as chosen by the user) is also shown. These reports provide both details
regarding the A/hr and ODI status of the patient and a general picture which can help the caregiver spot patient
trends in these areas.
This report is available only for adult patients, since A/hr and ODI are available only for adult patients.
In situations in which part of the information for the set period is missing (for example, SpO
2
was not measured
during a particular half hour during the 12 hours seen on the screen, so ODI could not be measured for that half
hour) the relevant time section will be shaded in grey, to indicate the presence of missing data. If the first hour
of monitoring has not yet been completed, the relevant time section on this screen will also be shaded in grey
(since A/hr and ODI are per-hour indicators, they cannot be calculated and thus cannot be displayed if less than
one hour of data is available).
Note:
This report is updated every 60 seconds. Since the A/hr and ODI numerics on the Home screen are
updated once a second, there may be a slight temporary discrepancy between the numerics on the two
screens from time to time.
Note:
The recommended use for this screen is that it be viewed when the caregiver wants to know more about
the patient’s apnea and O
2
desaturation events, or when an indicator on the Home screen indicates that
the report should be viewed. While the report is being displayed on the monitor screen, the display is not
updated, so it is best not to continually view this report on the monitor, but to continuously view the
Home screen while monitoring a patient. The EtCO
2
, FiCO
2
, RR, SpO
2
, and PR numerics are all updated
continuously on all screens, including the A/hr and O2 Desat Report screen.
The Apnea and O
2
Desat Report screen is reached with the
REPORTS
button on the menu on the Home screen.
The Print Report seen on the Print Report screen may be printed from the Apnea and O
2
Desat Report screen
with the built-in Capnostream printer (using the
START PRINTER
button), or saved to a USB flash memory
drive (using the
START USB
button).
Soft buttons on the Apnea and O
2
Desat Report screen are as follows:
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......