Demo Mode
Portable Bedside Capnograph/Pulse Oximeter
49
Note:
While the pump is off, CO
2
is not monitored and no breath waveform, etCO
2
, FiCO
2
, or respiration rate
number values are displayed. SpO
2
and pulse rate monitoring continues.
2. Pump Off mode can be ended by pressing the Pump Off button again.
3. Pump Off mode can be extended by using the control knob to select the
EXTEND TIMER
menu option.
When the monitor is in Pump Off mode, a timer appears in the message area at the top of the screen and shows
the total hours and minutes that the CO
2
monitoring has been turned off.
When the timer finishes or you manually exit Pump Off by pressing the Pump Off button again, the pump will
turn on and CO
2
monitoring will resume. The monitor automatically returns to the Home screen.
Note:
The Pump Off button does not function while scrolling in Graphical and Tabular Trend screens.
Demo Mode
The Capnostream monitor provides the possibility of viewing pre-recorded standard data to display an example
of the monitor's appearance under standard measurement conditions. The Demo Mode permits clinicians and
technicians to understand what the screen will show when monitoring patients, and is useful to them as a guide
before attaching the monitor to real patients.
T
O
U
SE THE
D
EMO
M
ODE
:
1. To enter the Demo Mode, click
SYSTEM
and then
SERVICE
on the menu bar at the bottom of the screen.
Enter the Service password (see
Changing Institutional Defaults
) to enter the Service screen.
2. On the Service screen, click the
DEMO MODE
button on the menu bar. The monitor will now enter Demo
Mode and will display pre-recorded CO
2
and SpO
2
data. As an indicator of operation under Demo mode, the
header will indicate
DEMO MODE
on a flashing orange background. An advisory message
DEMO MODE
- PRERECORDED DATA
will also appear. Service options and the Calibration Check feature are not
available to the user while the device is in Demo mode.
3. To exit the Demo Mode, you must turn off the monitor using the ON/OFF button at the front of the monitor.
When the monitor is turned on again, it will have returned to its standard operating status.
Monitor Screen Menu Reference Chart
The chart below shows the menu flow paths for navigating through the different screens of the Capnostream.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......