Changing Monitor Settings
Portable Bedside Capnograph/Pulse Oximeter
121
Figure 47 - Software Support Screen
Changing Monitor Settings
Alarm Limits
There are two sets of alarm limits that are stored in the monitor, for Adult/Pediatric and for Infant/Neonatal
patient types. (The Adult/Pediatric limits are relevant for adult patients and for all three pediatric patient types.)
The factory default settings for Adult and Neonatal alarm limits are given above in
To change the alarm limits for each type of patient, open the Institutional Defaults screen. Select
ALARM
LIMITS
and then use the control knob to select
SET INFANT/NEONATAL
or
SET ADULT/PEDIATRIC
or
SET ALARM DELAY
as required. You will then be presented with the Institutional Defaults: Alarm Limits:
Infant/Neonatal or the Institutional Defaults: Alarm Limits: Adult/Pediatric screen or the Alarm Delay screen.
To learn more about Alarm Delay options, see
Change the settings as described in the section
on page
By default, the Caution alarms are disabled, and in the Alarm Limits screen the numbers for the Caution alarm
settings appear in grey. Select
CAUTION ENABLE
if you want to enable the Caution alarms. When the
Caution alarms are enabled, the numbers for the Caution alarm settings change to white from grey.
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......