Warnings
Portable Bedside Capnograph/Pulse Oximeter
13
WARNING:
The FilterLine may ignite in the presence of O
2
when directly exposed to laser, ESU
devices, or high heat. When performing head and neck procedures involving laser,
electrosurgical devices or high heat, use with caution to prevent flammability of the
FilterLine or surrounding surgical drapes.
Electrical
WARNING:
To protect against electric shock hazard, the monitor’s cover is to be removed only by
qualified service personnel. There are no user-serviceable parts inside.
WARNING:
To ensure patient electrical isolation, connect only to other equipment with circuits that are
electrically isolated.
WARNING:
Connect the device only to a three-wire, grounded, hospital grade receptacle. The three-
conductor plug must be inserted into a properly wired three-wire receptacle; if a three-wire
receptacle is not available, a qualified electrician must install one in accordance with the
governing electrical code. Do not under any circumstances remove the grounding connector
from the power plug. Do not use extension cords or adapters of any type. The power cord
and plug must be intact and undamaged. To avoid the risk of electric shock, this equipment
must only be connected to a supply mains with protective earth.
WARNING:
Ensure that the monitor is positioned so that its mains plug is accessible for immediate
disconnection from supply mains, when needed.
WARNING:
If there is any doubt about the integrity of the protective earth conductor arrangement,
operate the device on internal battery power until the AC power supply protective
conductor is fully functional.
WARNING:
Do not connect to an electrical outlet controlled by a wall switch or a dimmer.
WARNING:
Measure the device's leakage current whenever an external device is connected to the serial
port. Leakage current must not exceed 100 microamperes.
WARNING:
To avoid risk of electric shock, this equipment must only be connected to a supply mains
with protective grounding.
WARNING:
Whenever the equipotential ground at the back of the device (reference
) is to be used, the user must connect to the pin in a way
which will ensure that accidental disconnection is avoided.
WARNING:
In a facility which provides detachable potential equalization conductors, the Equipotential
ground at the back of the device (reference
Figure 7 - Capnostream Rear View
may be used for optional connection between the Capnostream and the potential
equalization busbar of the electrical installation. The Equipotential ground at the back of the
device should not be used for a protective earth connection.
WARNING:
Always connect power cord to the device first, and then plug the power cord into the wall
outlet.
CAUTION:
Electrical installation of the room or the building in which the monitor is to be used must
comply with regulations specified by the country in which the equipment is to be used.
CAUTION:
Keep power cord, plug and socket clear in case an urgent power supply disconnection is
required.
Electro-magnetic Interference
This device has been tested and found to comply with the requirements for medical devices according to the
standard EN60601-1-2. This standard is designed to provide reasonable protection against harmful interference
in a typical medical installation.
However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical
noise in healthcare environments (for example: cellular phones, mobile two–way radios, electrical appliances),
Содержание Capnostream 20p
Страница 1: ...Capnostream 20p Portable Bedside Monitor Capnograph Pulse Oximeter Operator s Manual PN 012194C 0482...
Страница 20: ......
Страница 44: ......
Страница 54: ......
Страница 70: ......
Страница 94: ......
Страница 98: ...Printed Report Options 96 Portable Bedside Capnograph Pulse Oximeter Figure 36 Apnea and Desat Printed Report...
Страница 138: ......