Maintenance
3500/42 Operation and Maintenance
176
Full Scale Formulas - No Integration
Units
To Input RMS Volts
To Input Peak to Peak
Volts
g peak
(T.S.F. x Full-scale) x 0.707
(T.S.F. x Full-scale) x 2
g rms
(T.S.F x Full-scale)
(T.S.F. x Full-scale) x 2.82
m/s
2
peak
(T.S.F. x Full-scale) x 0.707
(T.S.F. x Full-scale) x 2
m/s
2
rms
(T.S.F x Full-scale)
(T.S.F. x Full-scale) x 2.82
T.S.F = Transducer Scale Factor. To use the formulas, the T.S.F. should be in
volts and the T.S.F. and full-scale values should both be of the same unit
system (metric or English). The transducer Scale Factor will always be
specified as volts per g pk or volts per m/s
2
pk.
Example 1:
Transducer Scale Factor = 100 mV/g
Full Scale = 2 g peak
For Peak to Peak input:
(0.100
×
2)
×
2 = 0.4 Vpp
For Vrms input:
(0.100
×
2)
×
0.707 = 0.1414 Vrms
Example 2:
Transducer Scale Factor = 10.19 mV/( m/s
2
)
Full Scale = 20 m/s
2
pk
For Peak to Peak input:
(0.01019
×
20)
×
2 = 0.4076 Vpp
For RMS input:
(0.01019
×
20)
×
0.707 = 0.1440 Vrms