Maintenance
3500/42 Operation and Maintenance
150
Test Equipment Setup - Velomitor (TMR I/O)
Simulate the transducer signal by connecting the power supply, function
generator, and multimeter to COM / A and SIG / B of channel 1 with polarity as
shown in the figure on page 153 (Velomitor Test Setup TMR I/O). Set the test
equipment as specified below.
Function Generator
Waveform: sinewave
DC Volts: 0 Vdc
Frequency: 100 Hz
Amplitude level: Minimum (above zero)
Seismoprobe Test Setup
The Test Equipment outputs should be floating relative to earth ground.
Multimeter
Function Generator
2.49 k
Ω
Power Supply