3500/42 Operation and Maintenance
Maintenance
97
Application Alert
Disconnecting the field
wiring will cause a not
OK condition.
Test Equipment Setup - Radial Vibration
Simulate the transducer signal by connecting the power supply, function
generator, and multimeter to COM and SIG of channel 1 with polarity as shown
in the figure below (Radial Vibration Test Setup). Set the test equipment as
specified below.
Power Supply
Function Generator
Keyphasor
Multiplier/Divider
-7.00 Vdc
Waveform: sinewave
DC Volts: 0 Vdc
Frequency: 100 Hz
Amplitude level:
Minimum (above zero)
Multiply Switch: 001
Divide Switch: 001